CPAP therapy
  in the Newborn
      Thrathip Kolatat M.D.
   Neonatal Intensive Care Unit
     Department of Pediatrics
Faculty of Medicine Siriraj Hospital
Continuous
 Positive
 Airway
 Pressure
 (CPAP)
CPAP
l   the application of positive
    airway pressure
    throughout the respiratory
    cycle during spontaneous
    respiration
l   history
    l   Harrison 1968: described
        grunting in neonates as
        naturally producing end-
                             end-
        expiratory pressure
    l   Gregory et al, 1971:
        introduced the clinical use
        of distending pressure in
        neonates
Physiologic effects of CPAP
l pulmonary mechanics
l cardiovascular stability
l pulmonary vascular resistance
Pulmonary effects
l decrease respiratory rate, tidal volume
  and minute volume
l regularization of respiration
l increase FRC and thoracic gas volume
l decrease lung compliance and
  dynamic compliance
l decrease total airway resistance
l protective effect on surfactant
Effect of CPAP on FRC in the infants with RDS

                   20

                   15


                   10
     FRC (ml/kg)



                                         Control
                    5
                                         +4 torr
                    0
                         2       3   4
                        Age (days)
Effect of different CPAP levels on PaO2
                 300




                 200
    Pa o(torr)
         2




                 100




                   0
                       4   8      12     16   20
                           CPAP (torr)
Cardiovascular effects
l   compromise venous return results in diminished cardiac
    output
    l   depend on lung compliance
    l   sign and symptom: tachycardia, metabolic acidosis,
        hypotension, decreased dynamic compliance, carbon dioxide
        retention

l   decrease peripheral and regional blood flow
l   decrease oxygen available to tissue
l   increase extra-pulmonary shunting secondary to an
    increase in pulmonary vascular resistance
Effects of CPAP
l   Renal function
    l   decrease renal blood flow
    l   decrease urine output and urinary sodium excretion
    l   increase antidiuretic hormone and aldosterone
l   Gastrointestinal function
    l   decrease gastrointestinal blood flow
    l   abdominal distention (CPAP belly syndrome)
l   Intracranial pressure (head box CPAP)
    l   increase intracranial pressure
    l   intracranial bleeding
Pressure volume curve
is divided into 3 regions
 l   region A: low lung                                                  Turbulent
                                                                        High-Velocity
                                                                                           Laminar
                                                                                         Low-Velocity
                                                                        Flow Region      Flow Region
     volume, low compliance
     and high resistance. The                                 .08                                        10,000




                                                                                                                  Airway Cross-Section in cm2
                                                                                Normal




                                Resistance (cm H2O/L(ml) s)
     P/V slope is low                                                          Newborn
                                                                                                         1000




                                             Volume per
                                                              .06
                                                                                 Lung


 l   region B: optimal lung                                   .04                                        100


     volume and increases                                     .02                                        10
                                                                                                     HMD Lung
     lung compliance                                                                                     2


 l   region C: high lung                                            0      5      10     15     20
     volume, low lung                                                       Airway Generation2O)
                                                                               Pressure (cmH

     compliance
                                Pressure- volume curve
Type
 of
CPAP
Clinical applications of CDP or CPAP
  l   respiratory distress syndrome
  l   meconium aspiration syndrome
  l   apnea of prematurity
  l   postoperative thoracotomy
  l   patent ductus arteriosus
  l   postoperative celosomia
  l   weaning patients from primary lung disease
  l   differentiation of primary lung disease from primary
      cardiac disease
  l   as adjunct to intermittent positive ventilation
  l   sleep apnea
  l   bronchomalacia
Clinical applications of CPAP
                           Low (2-3 cm H2O)

               Use                             Side effects
l maintenance of lung volume in      l may be too low to maintain
VLBW infants                         adequate lung volume or
l during weaning                     adequate oxygenation
l during hyperventilation in
PPHN
                         Medium (4-7 cm H2O)
               Use                             Side effects
l increasing lung volume in          If lungs have normal compliance
surfactant deficiency                l over distention
l stabilizing areas of atelectasis   l impede venous return
l stabilizing obstructed airway      l air leak
High (8-10 cm H2O)
              Use                             Side effects
l preventing alveolar collapse     l air leak
with poor CL and poor lung         l decreased CL if over distended
volume                             l may impede venous return
l improving distribution of
                                   l may increase PVR
ventilation
                                   l CO2 retention

                      Ultrahigh (11-15 cm H2O)
              Use                             Side effects
l tracheal or bronchial collapse   l air leak
l markedly decreased CL or         l decreased CL if over distended
severe obstruction                 l may impede venous return
l preventing white-out or re-      l may increase PVR
establishing lung volume during    l CO2 retention
ECHO
Tracheomalacia
Optimal CPAP level
     Disease                  Physiology              Optimal CPAP level

Acute RDS            l significant A-a gradient   l PaO2
                     l rapid change compliance    l oxygen consumption
                                                  l oxygen delivery

Small premature      l small A-a gradient         l   lung compliance
infant weaning off   l weak respiratory muscles
CPAP                 l increased chest wall
                     compliance
BPD on 60%           l decreased compliance       l pulmonary mechanics:
oxygen               l increased resistance       resistance, compliance

4-day-old RDS        l congestive heart failure   l balance between
and PDA              and pulmonary edema          cardiac output and
                                                  pulmonary blood flow
Clinical use of CPAP

l   clinical indications       l   follow up and
    l   sign of atelectasis on     weaning
        chest film                 l   follow-up PaO2 within
    l   chest wall retraction          15-20 min.
    l   require FiO2>0.5           l   weaning
    l   display rapidly                 l after oxygenation
        progressive lung disease           was improved
l   initial pressure setting            l extubation from

    l   nasal or nasopharyngeal:           CPAP 3-4 cm H2O
        6 cm H2O
    l   endotracheal: 4 cm H2O
Nursing care of CPAP
l   method                 l   nursing care
     l nasal                    l continuous care (oxygen,

     l nasopharyngeal             pressure)
     l endotracheal tube        l care of airway

     l face mask                     l prevent obstruction

l   components                       l prevent irritation of

     l oxygen                        nares
     l temperature and          l skin care

       humidity                 l abdominal distention

     l pressure                 l NPO
Indication of CPAP
Atelectatic disorders
l   PaO2 below 50-60 mm Hg in FiO2>0.6
l   recurrent apnea
Initial setup
l   CPAP 6 cm H2O
l   increase 2-cm. increments q 15 min. to a max.
    of 10 cm H2O or 12 cm H2O
l   increase FiO2 0.05-0.10 if PaO2<50 mm H2O
Respiratory distress syndrome




              l   improve survival rate,
                  especially larger infants
              l   modify course of the disease
              l   lower max.FiO2 required,
                  reduce total amount of time
                  under O2 and the need for
                  mechanical ventilation
Early CPAP in RDS
l   was proved to be more beneficial in the
    atelectatic disease
l   lower peak pressure required in infants treated
    with CPAP
l   enhance surfactant conservation
l   reduce the need for IMV by 20%, except infants
    with birth weight <1500 g.
l   improve mortality and decrease the incidence
    of BPD
l   prevent need for prolong intubation which
    reduce the incidence of acquired subglottic
    stenosis
Failure of CPAP therapy in RDS
l very low birth weight infant
l late application of CPAP
l severity of RDS
l associated disease e.g. sepsis,
  hypotension
l infants with severe degree of extra-
  pulmonary shunt
  (Fox and coworkers, 1977)
CPAP in apnea of prematurity
l   the application of low-level CPAP decrease
    the incidence of apnea of prematurity
    (compared to other forms of stimulation)
    l   improve oxygenation
    l   stimulation or inhibition of pulmonary reflexes
    l   alveolar stabilization
    l   mechanical splinting of airway; reduce
        supraglottic resistance in both inspiration and
        expiration
l   some investigators recommended the early
    use of CPAP as a preventive measure of
    apnea of prematurity
CPAP in an infant with MAS
l   pathology of meconium
    aspiration
     l atelectasis
     l large airway obstruction
     l V/Q abnormalities
l   application of low-to moderate
    level CPAP
     l resolution of atelectasis
     l stabilization of terminal
        airway
l   incidence of pneumothorax:
    not increased
l   precautions in case with PPHN
Adverse effects of CPAP
l   pulmonary air leaks
    l type of CPAP
    l lung compliance
    l gestational age

l gastric dilation and rupture
l hypotension
l increase pulmonary vascular resistance
l chronic lung disease ?
Complication of face mask CPAP

Más contenido relacionado

La actualidad más candente

Basic modes of mechanical ventilation
Basic modes of mechanical ventilationBasic modes of mechanical ventilation
Basic modes of mechanical ventilationLokesh Tiwari
 
Neonatal mechanical ventilation
Neonatal mechanical ventilationNeonatal mechanical ventilation
Neonatal mechanical ventilationChandan Gowda
 
18 basics of pediatric airway anatomy, physiology and management
18 basics of pediatric airway anatomy, physiology and management18 basics of pediatric airway anatomy, physiology and management
18 basics of pediatric airway anatomy, physiology and managementDang Thanh Tuan
 
Mechanical ventilation in neonates
Mechanical ventilation in neonatesMechanical ventilation in neonates
Mechanical ventilation in neonatespalpeds
 
Non Invasive Ventilation
Non Invasive VentilationNon Invasive Ventilation
Non Invasive VentilationManoj Prabhakar
 
Non invasive ventilation (niv)
Non invasive ventilation (niv)Non invasive ventilation (niv)
Non invasive ventilation (niv)Khairunnisa Azman
 
Continuous possitive pressure ventilation (cpap)
Continuous possitive pressure ventilation (cpap)Continuous possitive pressure ventilation (cpap)
Continuous possitive pressure ventilation (cpap)Malith Parakrama
 
Non invasive ventilation
Non invasive ventilationNon invasive ventilation
Non invasive ventilationtbf413
 
modes of ventilation
modes of ventilationmodes of ventilation
modes of ventilationNikhil Yadav
 
Non invasive ventilations
Non invasive ventilationsNon invasive ventilations
Non invasive ventilationsKIMS
 
Tracheostomy tubes by Dr.Ashwin Menon
Tracheostomy tubes by Dr.Ashwin MenonTracheostomy tubes by Dr.Ashwin Menon
Tracheostomy tubes by Dr.Ashwin MenonDr.Ashwin Menon
 
Manual respiratory bypass
Manual respiratory bypassManual respiratory bypass
Manual respiratory bypassMahesh kumar
 

La actualidad más candente (20)

Basic modes of mechanical ventilation
Basic modes of mechanical ventilationBasic modes of mechanical ventilation
Basic modes of mechanical ventilation
 
Oxygen delivery devices
Oxygen delivery devicesOxygen delivery devices
Oxygen delivery devices
 
Neonatal mechanical ventilation
Neonatal mechanical ventilationNeonatal mechanical ventilation
Neonatal mechanical ventilation
 
oxygen delivery devices -picu
oxygen delivery devices -picuoxygen delivery devices -picu
oxygen delivery devices -picu
 
Neonatal Ventilation
Neonatal VentilationNeonatal Ventilation
Neonatal Ventilation
 
Oxygen Therapy
Oxygen TherapyOxygen Therapy
Oxygen Therapy
 
18 basics of pediatric airway anatomy, physiology and management
18 basics of pediatric airway anatomy, physiology and management18 basics of pediatric airway anatomy, physiology and management
18 basics of pediatric airway anatomy, physiology and management
 
Mechanical ventilation in neonates
Mechanical ventilation in neonatesMechanical ventilation in neonates
Mechanical ventilation in neonates
 
Non Invasive Ventilation
Non Invasive VentilationNon Invasive Ventilation
Non Invasive Ventilation
 
Non invasive ventilation (niv)
Non invasive ventilation (niv)Non invasive ventilation (niv)
Non invasive ventilation (niv)
 
Continuous possitive pressure ventilation (cpap)
Continuous possitive pressure ventilation (cpap)Continuous possitive pressure ventilation (cpap)
Continuous possitive pressure ventilation (cpap)
 
Non invasive ventilation
Non invasive ventilationNon invasive ventilation
Non invasive ventilation
 
modes of ventilation
modes of ventilationmodes of ventilation
modes of ventilation
 
Non invasive ventilations
Non invasive ventilationsNon invasive ventilations
Non invasive ventilations
 
Tracheostomy tubes by Dr.Ashwin Menon
Tracheostomy tubes by Dr.Ashwin MenonTracheostomy tubes by Dr.Ashwin Menon
Tracheostomy tubes by Dr.Ashwin Menon
 
Manual respiratory bypass
Manual respiratory bypassManual respiratory bypass
Manual respiratory bypass
 
Prone Position
Prone PositionProne Position
Prone Position
 
NIPV for Peds
NIPV for PedsNIPV for Peds
NIPV for Peds
 
Humidfied nasal cannula
Humidfied nasal cannulaHumidfied nasal cannula
Humidfied nasal cannula
 
PRVC
PRVCPRVC
PRVC
 

Destacado

CPAP NASAL
CPAP NASALCPAP NASAL
CPAP NASALneoucin
 
NIV updated
NIV updatedNIV updated
NIV updatedEM OMSB
 
CPAP Machine Reviews 2013/14
CPAP Machine Reviews 2013/14CPAP Machine Reviews 2013/14
CPAP Machine Reviews 2013/141800CPAP.com
 
EMS CPAP Training
EMS CPAP TrainingEMS CPAP Training
EMS CPAP TrainingRobert Cole
 
Cpap final 2012_12 (2)
Cpap final 2012_12 (2)Cpap final 2012_12 (2)
Cpap final 2012_12 (2)dinterlandi
 
Asistencia ventilatoria
Asistencia ventilatoriaAsistencia ventilatoria
Asistencia ventilatoriaRafael Souza
 
Shop DRE's Complete Patient Monitor Series
Shop DRE's Complete Patient Monitor SeriesShop DRE's Complete Patient Monitor Series
Shop DRE's Complete Patient Monitor SeriesDRE Medical
 
iap-ahd-ventilation
iap-ahd-ventilationiap-ahd-ventilation
iap-ahd-ventilationNilu Panch
 
Pulse oximeter – how does it work and how can it be used in home care?
Pulse oximeter – how does it work and how can it be used in home care?Pulse oximeter – how does it work and how can it be used in home care?
Pulse oximeter – how does it work and how can it be used in home care?Silverline Meditech
 
Current Resp Trends In Neonatology
Current Resp Trends In NeonatologyCurrent Resp Trends In Neonatology
Current Resp Trends In NeonatologyDang Thanh Tuan
 
Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )
Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )
Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )Victoria Young
 
Noninvasive ventilation in neonates
Noninvasive ventilation in neonatesNoninvasive ventilation in neonates
Noninvasive ventilation in neonatesAziza Alamri - UOD
 
Danger signs in newborns
Danger signs in newbornsDanger signs in newborns
Danger signs in newbornsVarsha Shah
 
Non-Invasive Ventilation for Preterm Infants
Non-Invasive Ventilation for Preterm InfantsNon-Invasive Ventilation for Preterm Infants
Non-Invasive Ventilation for Preterm InfantsMark Weems
 

Destacado (20)

CPAP NASAL
CPAP NASALCPAP NASAL
CPAP NASAL
 
Cpap slides
Cpap slidesCpap slides
Cpap slides
 
NIV updated
NIV updatedNIV updated
NIV updated
 
CPAP Machine Reviews 2013/14
CPAP Machine Reviews 2013/14CPAP Machine Reviews 2013/14
CPAP Machine Reviews 2013/14
 
EMS CPAP Training
EMS CPAP TrainingEMS CPAP Training
EMS CPAP Training
 
Cpap final 2012_12 (2)
Cpap final 2012_12 (2)Cpap final 2012_12 (2)
Cpap final 2012_12 (2)
 
Uso de cpapn en neonatos
Uso de cpapn en neonatosUso de cpapn en neonatos
Uso de cpapn en neonatos
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
 
Asistencia ventilatoria
Asistencia ventilatoriaAsistencia ventilatoria
Asistencia ventilatoria
 
Shop DRE's Complete Patient Monitor Series
Shop DRE's Complete Patient Monitor SeriesShop DRE's Complete Patient Monitor Series
Shop DRE's Complete Patient Monitor Series
 
iap-ahd-ventilation
iap-ahd-ventilationiap-ahd-ventilation
iap-ahd-ventilation
 
Pulse oximeter – how does it work and how can it be used in home care?
Pulse oximeter – how does it work and how can it be used in home care?Pulse oximeter – how does it work and how can it be used in home care?
Pulse oximeter – how does it work and how can it be used in home care?
 
Current Resp Trends In Neonatology
Current Resp Trends In NeonatologyCurrent Resp Trends In Neonatology
Current Resp Trends In Neonatology
 
Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )
Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )
Design That Matters: Infant Pulse Oximeter ( MIT + RISD Collaboration )
 
Noninvasive ventilation in neonates
Noninvasive ventilation in neonatesNoninvasive ventilation in neonates
Noninvasive ventilation in neonates
 
Asthma management
Asthma managementAsthma management
Asthma management
 
Danger signs in newborns
Danger signs in newbornsDanger signs in newborns
Danger signs in newborns
 
Neonatal Emergencies
Neonatal EmergenciesNeonatal Emergencies
Neonatal Emergencies
 
Non-Invasive Ventilation for Preterm Infants
Non-Invasive Ventilation for Preterm InfantsNon-Invasive Ventilation for Preterm Infants
Non-Invasive Ventilation for Preterm Infants
 
Neonatal Problems
Neonatal ProblemsNeonatal Problems
Neonatal Problems
 

Similar a Cpap

Similar a Cpap (20)

Nonconventional Modes of Ventilation - Desphande
Nonconventional Modes of Ventilation - DesphandeNonconventional Modes of Ventilation - Desphande
Nonconventional Modes of Ventilation - Desphande
 
Respiratory physiology
Respiratory physiologyRespiratory physiology
Respiratory physiology
 
ARDS management
ARDS managementARDS management
ARDS management
 
Ards management
Ards managementArds management
Ards management
 
acute respiratory distress syndrome
acute respiratory distress syndromeacute respiratory distress syndrome
acute respiratory distress syndrome
 
Oxygen cascade & therapy
Oxygen cascade & therapyOxygen cascade & therapy
Oxygen cascade & therapy
 
Compliance
ComplianceCompliance
Compliance
 
Part 3compliance 3
Part 3compliance 3Part 3compliance 3
Part 3compliance 3
 
Applied respiratory physiology for Anaesthesiologist.pptx
Applied respiratory physiology for Anaesthesiologist.pptxApplied respiratory physiology for Anaesthesiologist.pptx
Applied respiratory physiology for Anaesthesiologist.pptx
 
FISIOLOGIA II - Ventilacion Pulmonar
FISIOLOGIA II - Ventilacion PulmonarFISIOLOGIA II - Ventilacion Pulmonar
FISIOLOGIA II - Ventilacion Pulmonar
 
One lung ventilation
One lung ventilationOne lung ventilation
One lung ventilation
 
Ventilator Graphics
Ventilator GraphicsVentilator Graphics
Ventilator Graphics
 
Capnography monitoring-PPT.pptx
Capnography monitoring-PPT.pptxCapnography monitoring-PPT.pptx
Capnography monitoring-PPT.pptx
 
High frequency ventilation.ppt
High frequency ventilation.pptHigh frequency ventilation.ppt
High frequency ventilation.ppt
 
Ards
ArdsArds
Ards
 
Small airways 2
Small airways 2Small airways 2
Small airways 2
 
Resp.pptx
Resp.pptxResp.pptx
Resp.pptx
 
Ards management
Ards managementArds management
Ards management
 
Respiratory physiology
Respiratory physiologyRespiratory physiology
Respiratory physiology
 
Respiration powerpoint
Respiration powerpointRespiration powerpoint
Respiration powerpoint
 

Más de Loveis1able Khumpuangdee (20)

Rollup01
Rollup01Rollup01
Rollup01
 
Protec
ProtecProtec
Protec
 
Factsheet hfm
Factsheet hfmFactsheet hfm
Factsheet hfm
 
Factsheet
FactsheetFactsheet
Factsheet
 
Eidnotebook54
Eidnotebook54Eidnotebook54
Eidnotebook54
 
Data l3 148
Data l3 148Data l3 148
Data l3 148
 
Data l3 147
Data l3 147Data l3 147
Data l3 147
 
Data l3 127
Data l3 127Data l3 127
Data l3 127
 
Data l3 126
Data l3 126Data l3 126
Data l3 126
 
Data l3 113
Data l3 113Data l3 113
Data l3 113
 
Data l3 112
Data l3 112Data l3 112
Data l3 112
 
Data l3 92
Data l3 92Data l3 92
Data l3 92
 
Data l3 89
Data l3 89Data l3 89
Data l3 89
 
Data l2 80
Data l2 80Data l2 80
Data l2 80
 
Hfm reccomment10072555
Hfm reccomment10072555Hfm reccomment10072555
Hfm reccomment10072555
 
Hfm work2550
Hfm work2550Hfm work2550
Hfm work2550
 
Factsheet hfm
Factsheet hfmFactsheet hfm
Factsheet hfm
 
Publichealth
PublichealthPublichealth
Publichealth
 
แนวทางการดาเน ํ นงานป ิ องก ้ นควบค ั มการระบาดของโรคม ุ ือ เท้า ปาก สําหรบแพ...
แนวทางการดาเน ํ นงานป ิ องก ้ นควบค ั มการระบาดของโรคม ุ ือ เท้า ปาก สําหรบแพ...แนวทางการดาเน ํ นงานป ิ องก ้ นควบค ั มการระบาดของโรคม ุ ือ เท้า ปาก สําหรบแพ...
แนวทางการดาเน ํ นงานป ิ องก ้ นควบค ั มการระบาดของโรคม ุ ือ เท้า ปาก สําหรบแพ...
 
hand foot mouth
hand foot mouthhand foot mouth
hand foot mouth
 

Último

❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...Sheetaleventcompany
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Dipal Arora
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Janvi Singh
 
tongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacytongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacyDrMohamed Assadawy
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...Namrata Singh
 
💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...
💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...
💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...Sheetaleventcompany
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Sheetaleventcompany
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...TanyaAhuja34
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowtanudubay92
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxSwetaba Besh
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryJyoti singh
 
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...Sheetaleventcompany
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...gragneelam30
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...soniyagrag336
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxSwetaba Besh
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Sheetaleventcompany
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan 087776558899
 
Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...
Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...
Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...Sheetaleventcompany
 

Último (20)

❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
tongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacytongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacy
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
 
💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...
💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...
💰Call Girl In Bangalore☎️7304373326💰 Call Girl service in Bangalore☎️Bangalor...
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...
Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...
Cheap Rate Call Girls Bangalore {9179660964} ❤️VVIP BEBO Call Girls in Bangal...
 

Cpap

  • 1. CPAP therapy in the Newborn Thrathip Kolatat M.D. Neonatal Intensive Care Unit Department of Pediatrics Faculty of Medicine Siriraj Hospital
  • 2. Continuous Positive Airway Pressure (CPAP)
  • 3. CPAP l the application of positive airway pressure throughout the respiratory cycle during spontaneous respiration l history l Harrison 1968: described grunting in neonates as naturally producing end- end- expiratory pressure l Gregory et al, 1971: introduced the clinical use of distending pressure in neonates
  • 4. Physiologic effects of CPAP l pulmonary mechanics l cardiovascular stability l pulmonary vascular resistance
  • 5. Pulmonary effects l decrease respiratory rate, tidal volume and minute volume l regularization of respiration l increase FRC and thoracic gas volume l decrease lung compliance and dynamic compliance l decrease total airway resistance l protective effect on surfactant
  • 6. Effect of CPAP on FRC in the infants with RDS 20 15 10 FRC (ml/kg) Control 5 +4 torr 0 2 3 4 Age (days)
  • 7. Effect of different CPAP levels on PaO2 300 200 Pa o(torr) 2 100 0 4 8 12 16 20 CPAP (torr)
  • 8. Cardiovascular effects l compromise venous return results in diminished cardiac output l depend on lung compliance l sign and symptom: tachycardia, metabolic acidosis, hypotension, decreased dynamic compliance, carbon dioxide retention l decrease peripheral and regional blood flow l decrease oxygen available to tissue l increase extra-pulmonary shunting secondary to an increase in pulmonary vascular resistance
  • 9.
  • 10. Effects of CPAP l Renal function l decrease renal blood flow l decrease urine output and urinary sodium excretion l increase antidiuretic hormone and aldosterone l Gastrointestinal function l decrease gastrointestinal blood flow l abdominal distention (CPAP belly syndrome) l Intracranial pressure (head box CPAP) l increase intracranial pressure l intracranial bleeding
  • 11. Pressure volume curve is divided into 3 regions l region A: low lung Turbulent High-Velocity Laminar Low-Velocity Flow Region Flow Region volume, low compliance and high resistance. The .08 10,000 Airway Cross-Section in cm2 Normal Resistance (cm H2O/L(ml) s) P/V slope is low Newborn 1000 Volume per .06 Lung l region B: optimal lung .04 100 volume and increases .02 10 HMD Lung lung compliance 2 l region C: high lung 0 5 10 15 20 volume, low lung Airway Generation2O) Pressure (cmH compliance Pressure- volume curve
  • 13. Clinical applications of CDP or CPAP l respiratory distress syndrome l meconium aspiration syndrome l apnea of prematurity l postoperative thoracotomy l patent ductus arteriosus l postoperative celosomia l weaning patients from primary lung disease l differentiation of primary lung disease from primary cardiac disease l as adjunct to intermittent positive ventilation l sleep apnea l bronchomalacia
  • 14. Clinical applications of CPAP Low (2-3 cm H2O) Use Side effects l maintenance of lung volume in l may be too low to maintain VLBW infants adequate lung volume or l during weaning adequate oxygenation l during hyperventilation in PPHN Medium (4-7 cm H2O) Use Side effects l increasing lung volume in If lungs have normal compliance surfactant deficiency l over distention l stabilizing areas of atelectasis l impede venous return l stabilizing obstructed airway l air leak
  • 15. High (8-10 cm H2O) Use Side effects l preventing alveolar collapse l air leak with poor CL and poor lung l decreased CL if over distended volume l may impede venous return l improving distribution of l may increase PVR ventilation l CO2 retention Ultrahigh (11-15 cm H2O) Use Side effects l tracheal or bronchial collapse l air leak l markedly decreased CL or l decreased CL if over distended severe obstruction l may impede venous return l preventing white-out or re- l may increase PVR establishing lung volume during l CO2 retention ECHO
  • 17. Optimal CPAP level Disease Physiology Optimal CPAP level Acute RDS l significant A-a gradient l PaO2 l rapid change compliance l oxygen consumption l oxygen delivery Small premature l small A-a gradient l lung compliance infant weaning off l weak respiratory muscles CPAP l increased chest wall compliance BPD on 60% l decreased compliance l pulmonary mechanics: oxygen l increased resistance resistance, compliance 4-day-old RDS l congestive heart failure l balance between and PDA and pulmonary edema cardiac output and pulmonary blood flow
  • 18. Clinical use of CPAP l clinical indications l follow up and l sign of atelectasis on weaning chest film l follow-up PaO2 within l chest wall retraction 15-20 min. l require FiO2>0.5 l weaning l display rapidly l after oxygenation progressive lung disease was improved l initial pressure setting l extubation from l nasal or nasopharyngeal: CPAP 3-4 cm H2O 6 cm H2O l endotracheal: 4 cm H2O
  • 19. Nursing care of CPAP l method l nursing care l nasal l continuous care (oxygen, l nasopharyngeal pressure) l endotracheal tube l care of airway l face mask l prevent obstruction l components l prevent irritation of l oxygen nares l temperature and l skin care humidity l abdominal distention l pressure l NPO
  • 20. Indication of CPAP Atelectatic disorders l PaO2 below 50-60 mm Hg in FiO2>0.6 l recurrent apnea Initial setup l CPAP 6 cm H2O l increase 2-cm. increments q 15 min. to a max. of 10 cm H2O or 12 cm H2O l increase FiO2 0.05-0.10 if PaO2<50 mm H2O
  • 21. Respiratory distress syndrome l improve survival rate, especially larger infants l modify course of the disease l lower max.FiO2 required, reduce total amount of time under O2 and the need for mechanical ventilation
  • 22. Early CPAP in RDS l was proved to be more beneficial in the atelectatic disease l lower peak pressure required in infants treated with CPAP l enhance surfactant conservation l reduce the need for IMV by 20%, except infants with birth weight <1500 g. l improve mortality and decrease the incidence of BPD l prevent need for prolong intubation which reduce the incidence of acquired subglottic stenosis
  • 23. Failure of CPAP therapy in RDS l very low birth weight infant l late application of CPAP l severity of RDS l associated disease e.g. sepsis, hypotension l infants with severe degree of extra- pulmonary shunt (Fox and coworkers, 1977)
  • 24. CPAP in apnea of prematurity l the application of low-level CPAP decrease the incidence of apnea of prematurity (compared to other forms of stimulation) l improve oxygenation l stimulation or inhibition of pulmonary reflexes l alveolar stabilization l mechanical splinting of airway; reduce supraglottic resistance in both inspiration and expiration l some investigators recommended the early use of CPAP as a preventive measure of apnea of prematurity
  • 25. CPAP in an infant with MAS l pathology of meconium aspiration l atelectasis l large airway obstruction l V/Q abnormalities l application of low-to moderate level CPAP l resolution of atelectasis l stabilization of terminal airway l incidence of pneumothorax: not increased l precautions in case with PPHN
  • 26. Adverse effects of CPAP l pulmonary air leaks l type of CPAP l lung compliance l gestational age l gastric dilation and rupture l hypotension l increase pulmonary vascular resistance l chronic lung disease ?
  • 27. Complication of face mask CPAP