2. WHEN A PHOTON BEAM INTERACTS
WITH MATTER, SOME PHOTON GETS
ABSORBED, SOME DEFLECTED, AND REST
IS TRANSMITTED TO PRODUCE A
RADIOGRAPHIC IMAGE
THE PART WHICH IS DEFLECTED FROM ITS
ORIGINAL PATH TO A NEW DIRECTION IS
SCATTER RADIATION
INTRODUCTION
3. SCATTER RADIATION ONLY ADDS UP TO UNWANTED DENSITIES
(NOISE) IN FILM
THE END RESULT IS THAT THE RADIOGRAPHIC IMAGE IS OF A
POOR QUALITY WITH DECREASED CONTRAST
DISADVANTAGES
4.
5. INCREASE IN FIELD SIZE
INCREASE IN PART THICKNESS
INCREASE IN KVP
CAUSE OF INCREASE IN
SCATTER RADIATION
6. WHEN FIELD SIZE IS NARROW, ONLY A
SMALL VOLUME OF TISSUE IS
IRRADIATED AND SO IT GENERATES A
SMALL NUMBER OF SCATTER
RADIATION
WHEN THE FIELD SIZE INCREASES,
SCATTER RADIATION TOO INCREASES
INITIALLY THEN REACHES A PLATEAU
THIS SATURATION POINT FOR SCATTER
RADIATION OCCURS WITH A FIELD OF
APPROX. 30 * 30 CM
FIELD SIZE
7. AS THE PART THICKNESS
INCREASES SCATTER RADIATON
TOO INCREASES UPTO A CERATAIN
EXTENT AND THEN SATURATION
OCCURS
THIS SATURATION IS BECAUSE
PHOTONS ORIGINATING IN THE
UPPER LAYERS OF THE PATIENT DO
NOT HAVE SUFFICIENT ENERGY TO
REACH THE FILM
PATIENT THICKNESS
8. IN A LOW ENERGY RANGE-
PHOTOELECTRIC EFFECT
PREDOMINATES- LITTLE
SCATTER RADIATION IS
PRODUCED
AS ENERGY INCREASES
COMPTON EFFECT
PREDOMINATES- SCATTER
RADIATION INCREASES
KVP
11. FILTRATION IS A PROCESS OF SHAPING
THE X RAY BEAM TO INCREASE THE
PHOTONS USEFUL FOR IMAGING.
IT ALSO DECREASES THE PATIENT DOSE
AND FILTERS THE LOW ENERGY
PHOTONS IN ORDER TO IMPROVE THE
IMAGE CONTRAST.
FILTERS
12. DIAGNOSTIC X RAY BEAMS ARE POLYCHROMATIC.
HIGH ENERGY PHOTONS PENETRATE THE BODY PART TO
PRODUCE A RADIO GRAPHIC IMAGE WHILE LOW ENERGY
PHOTONS GETS ABSORBED IN THE BODY, THUS INCREASING
THE PATIENT RADIATION DOSE.
FILTRATION ABSORBS THIS LOW ENERGY PHOTONS AND
INCREASES THE IMAGE CONTRAST.
HOW FILTRATION WORKS
15. ABSORPTION OF LOW ENERGY X RAY PHOTONS BY THE X
RAY TUBE COMPONENTS ITSELF IS INHERENT FILTRATION.
MATERIALS: GLASS ENVELOPE, INSULATING OIL, WINDOW
IN TUBE HOUSING.
MEASURED IN : ALUMINIUM EQUIVALENT ( 0.5 TO 1 MM ).
DISADVANTAGE : INCREASES THE MEAN ENERGY,
DECREASE THE TISSUE CONTRAST.
BERYLLIUM IS USED AS A EXIT PORTAL IN SOME
CIRCUMSTANCES WHERE UNFILTERED RADIATION IS
DESIRABLE
INHERENT FILTRATION
16.
17. ADDED FILTRATION IS A RESULT OF ANY BEAM ABSORBER
THAT IS PLACED IN THE PATH OF THE XRAY BEAM
IDEAL FILTER DOES NOT EXIST
MATERIALS: ALUMINUM AND COPPER.
ALUMINIUM (13) ABSORBS LOW ENERGY PHOTONS AND
FACES THE PATIENT
COPPER (29) ABSORBS HIGH ENERGY PHOTONS AND FACES
XRAY TUBE
ADDED FILTRATION
18. MOST FILTRATION OCCURS IN COPPER
WHY ALUMINUM ? IT ABSORBS CHARACTERISTIC
RADIATION OF 8 KEV PRODUCED BY PHOTO ELECTRIC
EFFECT IN COPPER.
THICKNESS : 2 MM THICKNESS ABSORBS ALL PHOTONS WITH
ENERGY < 20 KEV.
AS FURTHER THICKNESS INCREASES EVEN HIGH ENERGY
PHOTONS ARE ABSORBED RESULTING IN DECREASED
INTENSITY AND MORE EXPOSURE TIME.
19.
20. THEY ARE USED AT
PLACES OF DIFFERING
BODY PART THICKNESS
IN ORDER TO PRODUCE
A RADIO GRAPHIC
IMAGE OF UNIFORM
DENSITY.
WEDGE FILTERS
21. THESE ARE METALS WITH HIGH ATOMIC NUMBER > 60
EXCEPT Mo.
THESE FILTERS USE THE K- EDGE ABSORPTION OF THE
ELEMENT.
USE : IT ALLOWS ONLY THE XRAY BEAM WITH SPECIFIC
ENERGY RANGE.
THERE IS PRODUCTION OF A NARROW SPECTRUM, WITH
REDUCTION OF BOTH LOW AND HIGH ENERGY PHOTONS.
HEAVY METAL FILTERS ( K
EDGE FILTERS)
22. USED IN MAMMOGRAPHY AS A TARGET
AND FILTER.
AS A TARGET IT PRODUCES 17.5 KEV K
ALPHA AND 19.6 KEV K BETA
CHARACTERISTIC RADIATION.
ALSO PRODUCES X RAY RADIATION WITH
ENERGIES MORE THAN 20 KEV, WHEN LOW
KVP ( 30 -40) TECHNIQUE IS USED.
TO ABSORB THIS HIGH ENERGY
RADIATION, MOLYBDENUM IS USED AS A
FILTER.
MOLYBDENUM
23. GRIDS CONSISTS OF A SERIES OF LEAD FOIL STRIPS
SEPARATED BY XRAY TRANSPARENT SPACERS
INVENTED BY DR. GUSTAVE BUCKY
THE INTERSPACES ARE FILLED WITH ALUMINUM OR SOME
ORGANIC COMPOUND.
GRID IS POSITIONED BETWEEN THE PATIENT AND THE
IMAGE RECEPTOR.
THE GRID ID DESIGNED TO TRANSMIT ONLY THOSE X RAYS
WHOSE DIRECTION IS ON STRAIGHT LINE FROM SOURCE TO
IR.
GRIDS
24.
25.
26. LINEAR GRID
CROSSED GRID
PARALLEL GRID
FOCUSED GRID
PATTERNS OF GRID
27. LEAD STRIPS ARE PARALLEL TO EACH
OTHER IN LONGITUDINAL AXIS
ADVANTAGE : THEY ALLOW US TO
ANGLE THE XRAY TUBE ALONG THE
LENGTH OF THE GRID WITHOUT THE
LOSS OF PRIMARY RADIATION.
LINEAR GRID
28. MADE OF 2 SUPER IMPOSED LINEAR
GRIDS THAT HAVE THE SAME FOCUSING
DISTANCE.
GRID RATIO OF CROSSED GRIS IS EQUAL
TO THE SUM OF 2 LINEAR GRIDS.
DISADVANTAGE : CANNOT BE USED
WITH OBLIGE TECHNIQUES REQUIRING
ANGULATION OF THE X RAY TUBE
CROSSED GRID
29. GRID MADE UP OF LEAD STRIPS THAT ARE
SLIGHTLY ANGLED SO THAT THEY FOCUS
IN SPACE
MAY BE EITHER LINEAR OR CROSSED
LINEAR GRIDS FOCUS AT CONVERGENT
LINE
CROSS GRIDS FOCUS AT CONVERGENT
POINT
FOCUSED GRID
30. LEAD STRIPS ARE PARALLEL WHEN VIEWED IN
CROSS SECTION
FOCUSED AT INFINITY
USED IN FLUOROSCOPY
PARALLEL GRID
32. NUMBER OF LEAD STRIPS PER
INCH OF THE GRID
CALCULATED BY ADDING THE
THICKNESS OF THE LEAD
STRIPS AND INTER SPACES AND
DIVIDING THIS SUM INTO 1
LINES PER INCH
34. MEASUREMENT OF PERCENTAGE OF PRIMARY RADIATION
TRANSMITTED THROUGH THE GRID.
Tp = lp/lp' x 100
PRIMARY TRANSMISSION
35. RATIO OF THE INCIDENT RADIATION ON THE
GRID TO THE TRANSMITTED RADIATION
PASSING THROUGH THE GRID
MEASURE OF THE GRIDS ABILITY TO ABSORB
SCATTER RADIATION
UNLIKE PRIMARY TRANSMISSION, BUCKY
FACTOR INDICATES THE ABSORPTION OF
BOTH PRIMARY AND SECONDARY RADIATION
BUCKY FACTOR
36. RATIO OF THE CONTRAST WITH A GRID TO THE CONTRAST
WITHOUT A GRID
MEASURE OF THE GRIDS ABILITY TO IMPROVE CONTRAST
IT DEPENDS ON
-KVP
-FIELD SIZE
-PHANTOM THICKNESS
CONTRAST
IMPROVEMENT FACTOR
37. LOSS OF PRIMARY RADIATION THAT OCCURS WHEN THE
IMAGES IF THE LEAD STRIPS ARE PROJECTED WIDER THAN
THEY WOULD BE WITH ORDINARY MAGNIFICATION
RESULT OF A POOR GEOMETRIC RELATIONSHIP BETWEEN
THE PRIMARY BEAM AND THE LEAD FOIL STRIPS OF THE
GRID
CUTOFF IS COMPLETE AND NO PRIMARY RADIATION
REACHES THE FILM WHEN THE PROJECTED IMAGES OF THE
LEAD STRIPS ARE THICKER THAN THE WIDTH OF THE
INTERSPACES
GRID CUT OFF
38. AMOUNT OF CUTOFF IS ALWAYS GREATEST WITH HIGH RATIO
GRIDS AND SHORT GRID FOCUS DISTANCES
SITUATIONS THAT PRODUCE GRID CUTOFF
FOCUSSED GRIDS USED UPSIDE DOWN
LATERAL DECENTERING(GRID ANGULATION)-OFF CENTER
FOCUS GRID DISTANCE DECENTERING- OFF FOCUS
COMBINED LATERAL AND FOCUS GRID DISTANCE
DECENTERING
39.
40. THIS OCCURS WHEN A FOCUSSED GRID IS PLACED UPSIDE
DOWN ON THE IMAGE RECEPTOR, RESULTING IN THE GRID
LINES GOING OPPOSITE THE ANGLE OF DIVERGENCE OF THE
XRAY BEAM
RADIOGRAPHICALLY, THERE IS SIGNIFICANT LOSS OF
EXPOSURE ALONG THE EDGES OF THE IMAGE
UPSIDE DOWN FOCUSSED
GRIDS
41. PHOTONS EASILY PASS THROUGH THE
CENTER OF THE GRID BECAUSE THE
LEAD LINES ARE PERPENDICULAR TO
THE IMAGE RECEPTOR SURFACE
THE TILTED STRIPS NEAR THE EDGES
OF THE GRID WILL ABSORB
PROGRESSIVELY MORE RADIATION
CAUSING SEVERE PERIPHERAL CUTOFF
HOWEVER WITH PARALLEL GRIDS
EITHER SIDE MAY FACE THE TUBE OR
FILM
42. IT OCCURS WHEN THE CENTRAL RAY OF XRAY BEAM
IS NOT ALIGNED FROM SIDE TO SIDE WITH THE
CENTER OF FOCUSSED GRID
ALL THE LEAD STRIPS CUTOFF THE SAME AMOUNT
OF PRIMARY RADIATION
HENCE THERE IS UNIFORM LOSS OF RADIATION
OVER THE ENTIRE SURFACE OF THE GRID,
PRODUCING A UNIFORMLY LIGHT RADIOGRAPH
LATERAL DECENTERING-
OFF CENTER
43. TARGET OF THE XRAY TUBE IS
CORRECTLY CENTERED TO THE GRID
BUT IS POSITIONED ABOVE OR BELOW
THE CONVERGENT LINE
FOCUS GRID DISTANCE
DECENTERING- OFF FOCUS
44. CAUSES AN UNEVEN EXPOSURE , RESULTING IN A FIL THAT
IS DARK ON ONE SIDE AND LIGHT ON THE OTHER SIDE
COMBINED OFF CENTER
AND OFF FOCUSSED
45. OFF LEVEL GRID CUTOFF RESULTS
WHEN THE XRAY BEAM IS ANGLED
ACROSS THE LEAD STRIPS
OCCUR FROM EITHER A TUBE OR
THE GRID BEING ANGLED
OFTEN SEEN WITH PORTABLE
RADIOGRRAPH OR HORIZONTAL
BEAM EXAMINATIONS
OCCURS ONLY WITH BOTH
FOCUSSED AND PARALLEL GRIDS
OFF LEVEL GRIDS
46. FOCUSSED GRIDS ARE USUALLY USED
PLACED IN HOLDING MECHANISM THAT
BEGINS MOVING JUST BEFORE XRAY EXPOSURE
AND CONTINUES MOVING AFTER END OF
EXPOSURE
MOVES 1-3 INCHES
MOTION BLURS OUT THE GRID STRIPS
MOVING GRIDS
47. SINGLE STROKE: GRID HAD TO BE COCKED WITH A SPRING
MECHANISM
RECIPROCATING GRID: A MOTOR DRIVES THE GRID BACK
AND FORTH DURING THE EXPOSURE
OSCILLATING GRID: MOVES IN A CIRCULAR MOTION
AROUND THE GRID FRAME. DELICATE , SPRING LIKE
DEVICES LOCATED IN THE FOUR CORNERS HOLD THE GRID
CENTERED WITHIN THE FRAME
TYPES OF MOVING GRIDS
48. ADVANTAGES:
NO GRID LINES ARE SEEN
MOTION BLUR IS
UNDETECTABLE
DISADVANTAGES:
COSTLY
MECHANICAL PROBLEMS
MAY OCCUR
INCREASE THE PATIENT
RADIATION DOSE