GROUP 7
        ADVANCES IN ION SELECTIVE
            ELECTRODES(ISE)


                GROUP MEMBERS:
YAAKOB BIN ABD RAZAK                  (154631)
AMIR BIN HASANUDIN                    (154046)
ASILAH BTE JAMIL                      (154423)
NOR HIDAYAT BINTI YUSOF               (152356)
INTAN NORYANA BINTI AHMAD             (153741)
NORSYAMIMI BINTI CHE SULAMAN          (153504)
NUR SHUHADA BINTI MOHD MOKHTARUDDIN   (153142)
NURUL FADZILLAH BINTI MOHD HATTA      (152266)
FAZURA EMYZA BINTI ABD AZIZ           (153819)
NUR FATIHAH BINTI ABAS                (154120)
Preview
1) Introduction to Ion Selective Electrodes,ISE

2) Composition of ISE(Glass Membrane Electrode,Solid State
Electrode,Liquid Membrane Electrode,Gas Sensing Electrode)

3) Parameters of ISE(sensitivity,selectivity,detection limit and
response time)

4) Potentionmetric measurements of ISE(in vivo, on line,on site)

5) The Recent advance(application of ISE)
Ion selective electrodes(ISE)
-An ion-selective electrode (ISE), also known as a specific ion
electrode (SIE), is a transducer (or sensor) that converts the activity of
a specific ion dissolved in a solution into an electrical potential, which
can be measured by a voltmeter or pH meter.

-Several types of sensing electrodes are commercially available which
are Glass Membrane Electrode, Solid State Electrode, Liquid
Membrane Electrode,Gas Sensing Electrode
• Ion Selective Electrodes (including the most common pH electrode) work on the
  basic principal of the galvanic cell .By measuring the electric potential generated
  across a membrane by "selected" ions, and comparing it to a reference
  electrode, a net charge is determined. The strength of this charge is directly
  proportional to the concentration of the selected ion. The basic formula is given for
  the galvanic cell:

                             Ecell = EISE - ERef
TYPES OF ION SELECTIVE
   ELECTRODE (ISE)

 •Glass Membrane Electrode
 •Solid State Electrode
 •Liquid Membrane Electrode
 •Gas Sensing Electrode
GLASS MEMBRANE ELECTRODE
• Glass electrode are responsive to univalent cations
  ( H+ , Na+)
• The selectivity for this cation by varying the composition
  of a thin ion sensitive glass membrane.
• Example: pH electrode
  - used for pH measurement
  - used as a transducer in various gas and biocatalytic
  sensor, involving proton generating or consuming
  reaction.
pH electrode




•Glass membrane manufactured from SiO2 with negatively
charged oxygen atom.
•Inside the glass bulb, a dilute HCl solution and silver wire
coated with a layer of silver chloride.
•The electrode is immersed in the solution and pH is
measured.
SOLID STATE ELECTRODE

 Solid state electrode are selective primarily to
  anions.
 It may be a homogenous membrane electrode or
  heterogeneous membrane electrode.
 Homogenous membrane electrode: ion-selective
  electrodes in which the membrane is a crystalline
  material (AgI/Ag2S).
 Heterogeneous membrane electrode: ion-selective
  electrodes prepared of an active substance, or mixture
  of active substances (silicone rubber or PVC).
 Example: Fluoride ion selective electrode
LIQUID MEMBRANE ELECTRODE
 Liquid membrane is a type of ISE based on water-
  immiscible liquid substances produced in a polymeric
  membrane used for direct potentiometric
  measurement.
 Used for direct measurement of several polyvalent
  cations (Ca ion) as well as a certain anions.
 Example: Ion Exchanger Electrode




 •The polymeric membrane made of PVC to separate the
 test solution from its inner compartment.
 •Contains standard solution of the target ion.
 •The filling solution contains a chloride salt for establishing
 the potential of the internal Ag/AgCl wire electrode.
GAS SENSING ELECTRODE
 Available for the measurement of ammonia, carbon
  dioxide and nitrogen oxide.
 This type of electrode consist of permeable
  membrane and an internal buffer solution.
 The pH of the buffer changes as the gas react with it.
 The change is detected by a combination pH sensor
  within the housing.
 This type of electrode does not require an external
  reference electrode.
Parameters of ISE
       Sensitivity
       Selectivity
     Detection Limit
     Response Time
Nikolsky-Eisenman equation,



Where
 RT.ln10/F-Sensitivity(S)
 kxy-selectivity coefficient
 ax-activity of the ion, X
 ay-activity of the interfering ion, Y
 zx-charge of the primary ion, X
 zy-charge on the interfering ion, Y
 The ‘constant’-E0
Selectivity


 An electrode used to measure primary ion X selectively
  may also slightly responds to interference ion Y.
 Selectivity coefficient, kxy,is used in ISE to distinguish the
  ion X from ion Y in the same solution.
 kxy is not constant and depends on several factors including
  the concentration of both elements, the total ionic strength
  of the solution, and the temperature.
 The value of kxy, is defined by the Nikolsky-Eisenman
  equation:
    The smaller the value of kxy ,the greater is the electrode’s
     preference for the primary ion, X.
Sensitivity


 Ability to detect primary ion at the lower concentration.
 Theoretical value of S=59mV,represents 100%
  sensitivity.
 The value of S varies with kxy
 small value of kxy means that the electrode is more
  sensitive to primary ion,X instead of interfering ion,Y.
Detection limit
 Defined by the intersection of the two extrapolated linear
  parts of the ion-selective calibration curve.
 LOD ~ 10-5-10-6 M is measured for most ISE.
 Observed LOD is often governed by the presence of other
  interfering ions or impurities.
 Metal buffers can be used to eliminate the effects which
  lead to the contamination of very dilute solutions. May
  reduce LOD to 10-10 M.
Response time
 From the time the ISE and a reference electrode are
  dipped in the sample solution (or the time at which the
  ion concentration in a solution in contact with ISE and a
  reference electrode is changed ) to when the potential of
  the cell becomes equal to its steady-state value within 1
  [mV] or has reached 90% of the final value (in certain
  cases also 63% or 95%).

 The response time usually increases with decreasing
  determinand concentration
Time constant of                                           Rate of the charge-
  the measuring              Determinand                     transfer reaction
    instrument          diffusion through the                    across the
                         hydrodynamic layer                 membrane/solution
                                                              interface, which
                                                           results in charging of
                                                           the electrical double
                                                          layer at this interface
Rate of the exchange
reaction between the               Factors
 determinand in the               influence
   membrane and an
  interferent in the              response
                                     time                  Dissolution of the
       analyte                                             membrane-active
                                                           component in the
                                                                analyte




           Establishment of the                 Interferent
            diffusion potential               diffusion in the
           across the membrane                   membrane
ON-LINE, ON SITE AND IN VIVO
POTENTIOMETRIC MEASUREMENTS
 Meanings:
     1) on-line : being in progress now.
     2) on-site : taking place or located at the side.
     3) in vivo : within a living organism.
   ISE have been widely used as detectors in high-speed automated flow
    analyzers such as air-segmented or flow injection systems for the high-speed
    determination of physiologically important cationic or anionic electrolytes in
    body fluids.



                                                         Example of flow injection
                                                      determination of physiologically
                                                           potassium in serum.




   Several designs of low-volume potentiometric flow detectors
have been reported.
                                                            Flow-through
                                                      potentiometric cell design
Potentiometric microelectrodes are very suitable for in vivo real time
clinical monitoring of blood electrolytes, intracellular studies, in situ
environmental surveillance or industrial process control. For example
Simon’s group described the utility of a system for on-line measurement
of blood potassium ion concentration during an open-heart surgery.

                                         Miniaturized catheter-type ISE sensors
                                         such as implantable probe represent
                                         the preferred approach for routine
                                         clinical in vivo monitoring of blood
                                         electrolyte.
                                         Diamond's group developed an array of
                                         miniaturized chloride, sodium, and
                                         potassium ISEs for point-of-care
                                         analysis of sweat in connection to non-
                                         invasive diagnosis of cystic fibrosis.



 Miniaturized catheter-type ISE sensor
Recent Advancement of Ion Selective
          Electrode (ISE)
Agriculture and Fishery                                Pollution Control
-Soil and fertilizer for                               -pH of acid rain, soil, surface water
Nitrate, Ammonium,                                     -Contamination of surface water
 Potassium to optimize the use of                       and ground water with ammonium
fertilizer.                                             and nitrate
-Dissolved Oxygen and pH in ponds                      -Contamination of waste water with
for                                                    Cyanide, Cadmium, Mercury and
 fish breeding.                                        Copper


Industrial Production
                                                         Food and Quality Control
-Salinity and pH of Boiler
feed
                                    Worldwide            -Nitrate and Nitrite in meat and
                                    application          vegetables
 water
                                                         -Chloride, Sodium, Nitrate and
-Cyanide in plating baths
                                                         Nitrite in
-Process (specific ions)
                                                          baby food.
                                                         -Cadmium in fish



               Medical Diagnosis and Hygiene Control
               -Potassium in urine
               -Contamination in various ions.
Advantages of Ion Selective Electrode
         (ISE) Technique
 When compared to many other analytical techniques, Ion-
  Selective Electrodes are relatively inexpensive and simple to
  use and have an extremely wide range of applications and wide
  concentration range.
 Under the most favourable conditions, when measuring ions in
  relatively dilute aqueous solutions and where interfering ions are
  not a problem, they can be used very rapidly and easily.
 They are particularly useful in applications where only an order
  of magnitude concentration is required, or it is only necessary to
  know that a particular ion is below a certain concentration level.
    They are invaluable for the continuous monitoring of changes in
    concentration for example in potentiometric titrations or monitoring
    the uptake of nutrients, or the consumption of reagents.
 They are particularly useful in biological/medical applications
    because they measure the activity of the ion directly, rather than the
    concentration.
   ISEs are one of the few techniques which can measure both
    positive and negative ions.
 They are unaffected by sample colour or turbidity.
    ISEs can be used in aqueous solutions over a wide temperature
    range. Crystal membranes can operate in the range 0 C to 80 C
    and plastic membranes from 0 C to 50 C.
Surface Plasmon Resonance (SPR)
   Based on immunosensors for detection and monitoring of low-molecular-
    weight analytes of biomedical, food and environmental fields.
   SPR is a surface sensitive optical technique for monitoring biomolecular
    interactions occurring in very close vicinity of a transducer (gold)
    surface, and that has given it a great potential for studying surface-confined
    affinity interactions without rinsing out unreacted or excess reactants in
    sample solutions.
   It allows real-time study of the binding interactions between a biomolecule
    (antibody) immobilized on a transducer surface with its biospecific partner
    (analyte) in solution without the need for labeling the biomolecules by
    exploiting the interfacial refractive index changes associated with any
    affinity binding interaction.
   In general, an SPR immunosensor is comprised of several important
    components: a light source, a detector, a transduction surface (usually
    gold-film), a prism, biomolecule (antibody or antigen) and a flow system.
• The transduction surface is usually a thin gold-film (50–100 nm) on a
glass slide optically coupled to a glass prism through a refractive index
matching oil. In addition to gold, several metals can be used including
silver, copper and aluminium. However, gold is highly preferred due to its
chemical stability and free electron behaviour.
Thank you for your attention

Más contenido relacionado

La actualidad más candente

Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.Sandhya Talla
 
Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..vishnupriya456
 
Ion selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterIon selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterMuhammad Asif Shaheeen
 
ION CHROMATOGRAPHY ION CHROMATOGRAPHY
ION CHROMATOGRAPHY ION CHROMATOGRAPHYION CHROMATOGRAPHY ION CHROMATOGRAPHY
ION CHROMATOGRAPHY ION CHROMATOGRAPHYShikha Popali
 
Polarography and Voltammetry
Polarography and VoltammetryPolarography and Voltammetry
Polarography and VoltammetrySudha durairaj
 
Coulometry.pptx presentation assignment copy
Coulometry.pptx presentation assignment   copyCoulometry.pptx presentation assignment   copy
Coulometry.pptx presentation assignment copyKibetDerrick
 
Stripping voltammetry
Stripping voltammetryStripping voltammetry
Stripping voltammetryRituHaldive
 
Mass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPUMass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPUsiraj174
 

La actualidad más candente (20)

Flame Atomic Absorption Spectroscopy
Flame Atomic Absorption SpectroscopyFlame Atomic Absorption Spectroscopy
Flame Atomic Absorption Spectroscopy
 
Coulometry
CoulometryCoulometry
Coulometry
 
1 Potentiometry
1  Potentiometry1  Potentiometry
1 Potentiometry
 
Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.
 
Cyclic voltammetry
Cyclic voltammetryCyclic voltammetry
Cyclic voltammetry
 
Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..Ion sensitive electrode & gas sensitive electrode..
Ion sensitive electrode & gas sensitive electrode..
 
coulorometry
coulorometrycoulorometry
coulorometry
 
Ion selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meterIon selective electrodes ISE (potentiometery) PH meter
Ion selective electrodes ISE (potentiometery) PH meter
 
ION CHROMATOGRAPHY ION CHROMATOGRAPHY
ION CHROMATOGRAPHY ION CHROMATOGRAPHYION CHROMATOGRAPHY ION CHROMATOGRAPHY
ION CHROMATOGRAPHY ION CHROMATOGRAPHY
 
Polarography and Voltammetry
Polarography and VoltammetryPolarography and Voltammetry
Polarography and Voltammetry
 
TYPES OF POLAROGRAPHY.pptx
TYPES OF POLAROGRAPHY.pptxTYPES OF POLAROGRAPHY.pptx
TYPES OF POLAROGRAPHY.pptx
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Coulometry.pptx presentation assignment copy
Coulometry.pptx presentation assignment   copyCoulometry.pptx presentation assignment   copy
Coulometry.pptx presentation assignment copy
 
Capillary electrophoresis
Capillary electrophoresisCapillary electrophoresis
Capillary electrophoresis
 
Stripping voltammetry
Stripping voltammetryStripping voltammetry
Stripping voltammetry
 
Mass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPUMass spectroscopy for M Sc I Chemistry SPPU
Mass spectroscopy for M Sc I Chemistry SPPU
 
Ion selective electrode (hw)85
Ion selective electrode (hw)85Ion selective electrode (hw)85
Ion selective electrode (hw)85
 
AMPEROMETRY
AMPEROMETRYAMPEROMETRY
AMPEROMETRY
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
AES Plasma
AES PlasmaAES Plasma
AES Plasma
 

Similar a Advances in Ion Selective Electrodes(ISE)

Biosensors
BiosensorsBiosensors
Biosensorsfirru
 
Biosensors of modern era
Biosensors of modern eraBiosensors of modern era
Biosensors of modern eraSrilaxmiMenon
 
Biosensor and its Applications
Biosensor and its ApplicationsBiosensor and its Applications
Biosensor and its ApplicationsJishnukc1
 
46
4646
46dL21
 
Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...
Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...
Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...Fawad Majeed
 
Electrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseElectrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseAngelica Nhoj Gemora
 
Electronic pill
Electronic pillElectronic pill
Electronic pillb4ujos
 
Excitable Cells: Revision Notes
Excitable Cells: Revision NotesExcitable Cells: Revision Notes
Excitable Cells: Revision Notesmeducationdotnet
 
intro to biosensor
intro to biosensorintro to biosensor
intro to biosensormanjurekar
 
Methods in Electrochemical in chemistry
Methods in Electrochemical  in chemistryMethods in Electrochemical  in chemistry
Methods in Electrochemical in chemistryKimEliakim1
 
biosensors;components,types , applications and GMO biosensors
biosensors;components,types , applications and GMO biosensorsbiosensors;components,types , applications and GMO biosensors
biosensors;components,types , applications and GMO biosensorsSilpa
 
application-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdf
application-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdfapplication-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdf
application-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdfLinhNguyenTien3
 

Similar a Advances in Ion Selective Electrodes(ISE) (20)

Biosensors
BiosensorsBiosensors
Biosensors
 
Biosensors of modern era
Biosensors of modern eraBiosensors of modern era
Biosensors of modern era
 
Biosensor and its Applications
Biosensor and its ApplicationsBiosensor and its Applications
Biosensor and its Applications
 
46
4646
46
 
Biosensor
BiosensorBiosensor
Biosensor
 
Biosensors
BiosensorsBiosensors
Biosensors
 
Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...
Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...
Nanowire Based FET Biosensors and Their Biomedical Applications. Fawad Majeed...
 
Biosensors
BiosensorsBiosensors
Biosensors
 
Electrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseElectrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, ise
 
Biosensors
BiosensorsBiosensors
Biosensors
 
Electronic pill
Electronic pillElectronic pill
Electronic pill
 
Excitable Cells: Revision Notes
Excitable Cells: Revision NotesExcitable Cells: Revision Notes
Excitable Cells: Revision Notes
 
intro to biosensor
intro to biosensorintro to biosensor
intro to biosensor
 
Potentiometry ppt By Chand.pptx
Potentiometry ppt By Chand.pptxPotentiometry ppt By Chand.pptx
Potentiometry ppt By Chand.pptx
 
Methods in Electrochemical in chemistry
Methods in Electrochemical  in chemistryMethods in Electrochemical  in chemistry
Methods in Electrochemical in chemistry
 
Electrophoresis
ElectrophoresisElectrophoresis
Electrophoresis
 
biosensors;components,types , applications and GMO biosensors
biosensors;components,types , applications and GMO biosensorsbiosensors;components,types , applications and GMO biosensors
biosensors;components,types , applications and GMO biosensors
 
application-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdf
application-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdfapplication-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdf
application-data-sheet-theory-application-of-conductivity-rosemount-en-68442.pdf
 
Biosensors
BiosensorsBiosensors
Biosensors
 
PAP
PAPPAP
PAP
 

Más de Nur Fatihah

INFRARED SPECTROSCOPY
INFRARED SPECTROSCOPYINFRARED SPECTROSCOPY
INFRARED SPECTROSCOPYNur Fatihah
 
Spectroscopy 5201
Spectroscopy 5201Spectroscopy 5201
Spectroscopy 5201Nur Fatihah
 
Polymer Chemistry
Polymer ChemistryPolymer Chemistry
Polymer ChemistryNur Fatihah
 
Viscosity measurement
Viscosity measurementViscosity measurement
Viscosity measurementNur Fatihah
 
Fatty acids and triglycerides
Fatty acids and triglyceridesFatty acids and triglycerides
Fatty acids and triglyceridesNur Fatihah
 
Chemistry of alkaloid
Chemistry of alkaloidChemistry of alkaloid
Chemistry of alkaloidNur Fatihah
 
Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...
Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...
Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...Nur Fatihah
 

Más de Nur Fatihah (7)

INFRARED SPECTROSCOPY
INFRARED SPECTROSCOPYINFRARED SPECTROSCOPY
INFRARED SPECTROSCOPY
 
Spectroscopy 5201
Spectroscopy 5201Spectroscopy 5201
Spectroscopy 5201
 
Polymer Chemistry
Polymer ChemistryPolymer Chemistry
Polymer Chemistry
 
Viscosity measurement
Viscosity measurementViscosity measurement
Viscosity measurement
 
Fatty acids and triglycerides
Fatty acids and triglyceridesFatty acids and triglycerides
Fatty acids and triglycerides
 
Chemistry of alkaloid
Chemistry of alkaloidChemistry of alkaloid
Chemistry of alkaloid
 
Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...
Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...
Au nanospheres and nanorods for enzyme-free electrochemical biosensor applica...
 

Advances in Ion Selective Electrodes(ISE)

  • 1. GROUP 7 ADVANCES IN ION SELECTIVE ELECTRODES(ISE) GROUP MEMBERS: YAAKOB BIN ABD RAZAK (154631) AMIR BIN HASANUDIN (154046) ASILAH BTE JAMIL (154423) NOR HIDAYAT BINTI YUSOF (152356) INTAN NORYANA BINTI AHMAD (153741) NORSYAMIMI BINTI CHE SULAMAN (153504) NUR SHUHADA BINTI MOHD MOKHTARUDDIN (153142) NURUL FADZILLAH BINTI MOHD HATTA (152266) FAZURA EMYZA BINTI ABD AZIZ (153819) NUR FATIHAH BINTI ABAS (154120)
  • 2. Preview 1) Introduction to Ion Selective Electrodes,ISE 2) Composition of ISE(Glass Membrane Electrode,Solid State Electrode,Liquid Membrane Electrode,Gas Sensing Electrode) 3) Parameters of ISE(sensitivity,selectivity,detection limit and response time) 4) Potentionmetric measurements of ISE(in vivo, on line,on site) 5) The Recent advance(application of ISE)
  • 3. Ion selective electrodes(ISE) -An ion-selective electrode (ISE), also known as a specific ion electrode (SIE), is a transducer (or sensor) that converts the activity of a specific ion dissolved in a solution into an electrical potential, which can be measured by a voltmeter or pH meter. -Several types of sensing electrodes are commercially available which are Glass Membrane Electrode, Solid State Electrode, Liquid Membrane Electrode,Gas Sensing Electrode
  • 4. • Ion Selective Electrodes (including the most common pH electrode) work on the basic principal of the galvanic cell .By measuring the electric potential generated across a membrane by "selected" ions, and comparing it to a reference electrode, a net charge is determined. The strength of this charge is directly proportional to the concentration of the selected ion. The basic formula is given for the galvanic cell: Ecell = EISE - ERef
  • 5. TYPES OF ION SELECTIVE ELECTRODE (ISE) •Glass Membrane Electrode •Solid State Electrode •Liquid Membrane Electrode •Gas Sensing Electrode
  • 6. GLASS MEMBRANE ELECTRODE • Glass electrode are responsive to univalent cations ( H+ , Na+) • The selectivity for this cation by varying the composition of a thin ion sensitive glass membrane. • Example: pH electrode - used for pH measurement - used as a transducer in various gas and biocatalytic sensor, involving proton generating or consuming reaction.
  • 7. pH electrode •Glass membrane manufactured from SiO2 with negatively charged oxygen atom. •Inside the glass bulb, a dilute HCl solution and silver wire coated with a layer of silver chloride. •The electrode is immersed in the solution and pH is measured.
  • 8. SOLID STATE ELECTRODE  Solid state electrode are selective primarily to anions.  It may be a homogenous membrane electrode or heterogeneous membrane electrode.  Homogenous membrane electrode: ion-selective electrodes in which the membrane is a crystalline material (AgI/Ag2S).
  • 9.  Heterogeneous membrane electrode: ion-selective electrodes prepared of an active substance, or mixture of active substances (silicone rubber or PVC).  Example: Fluoride ion selective electrode
  • 10. LIQUID MEMBRANE ELECTRODE  Liquid membrane is a type of ISE based on water- immiscible liquid substances produced in a polymeric membrane used for direct potentiometric measurement.  Used for direct measurement of several polyvalent cations (Ca ion) as well as a certain anions.
  • 11.  Example: Ion Exchanger Electrode •The polymeric membrane made of PVC to separate the test solution from its inner compartment. •Contains standard solution of the target ion. •The filling solution contains a chloride salt for establishing the potential of the internal Ag/AgCl wire electrode.
  • 12. GAS SENSING ELECTRODE  Available for the measurement of ammonia, carbon dioxide and nitrogen oxide.  This type of electrode consist of permeable membrane and an internal buffer solution.  The pH of the buffer changes as the gas react with it.
  • 13.  The change is detected by a combination pH sensor within the housing.  This type of electrode does not require an external reference electrode.
  • 14. Parameters of ISE  Sensitivity  Selectivity  Detection Limit  Response Time
  • 15. Nikolsky-Eisenman equation, Where  RT.ln10/F-Sensitivity(S)  kxy-selectivity coefficient  ax-activity of the ion, X  ay-activity of the interfering ion, Y  zx-charge of the primary ion, X  zy-charge on the interfering ion, Y  The ‘constant’-E0
  • 16. Selectivity  An electrode used to measure primary ion X selectively may also slightly responds to interference ion Y.  Selectivity coefficient, kxy,is used in ISE to distinguish the ion X from ion Y in the same solution.  kxy is not constant and depends on several factors including the concentration of both elements, the total ionic strength of the solution, and the temperature.  The value of kxy, is defined by the Nikolsky-Eisenman equation:  The smaller the value of kxy ,the greater is the electrode’s preference for the primary ion, X.
  • 17. Sensitivity  Ability to detect primary ion at the lower concentration.  Theoretical value of S=59mV,represents 100% sensitivity.  The value of S varies with kxy  small value of kxy means that the electrode is more sensitive to primary ion,X instead of interfering ion,Y.
  • 18. Detection limit  Defined by the intersection of the two extrapolated linear parts of the ion-selective calibration curve.  LOD ~ 10-5-10-6 M is measured for most ISE.  Observed LOD is often governed by the presence of other interfering ions or impurities.  Metal buffers can be used to eliminate the effects which lead to the contamination of very dilute solutions. May reduce LOD to 10-10 M.
  • 19.
  • 20. Response time  From the time the ISE and a reference electrode are dipped in the sample solution (or the time at which the ion concentration in a solution in contact with ISE and a reference electrode is changed ) to when the potential of the cell becomes equal to its steady-state value within 1 [mV] or has reached 90% of the final value (in certain cases also 63% or 95%).  The response time usually increases with decreasing determinand concentration
  • 21. Time constant of Rate of the charge- the measuring Determinand transfer reaction instrument diffusion through the across the hydrodynamic layer membrane/solution interface, which results in charging of the electrical double layer at this interface Rate of the exchange reaction between the Factors determinand in the influence membrane and an interferent in the response time Dissolution of the analyte membrane-active component in the analyte Establishment of the Interferent diffusion potential diffusion in the across the membrane membrane
  • 22. ON-LINE, ON SITE AND IN VIVO POTENTIOMETRIC MEASUREMENTS
  • 23.  Meanings: 1) on-line : being in progress now. 2) on-site : taking place or located at the side. 3) in vivo : within a living organism.
  • 24. ISE have been widely used as detectors in high-speed automated flow analyzers such as air-segmented or flow injection systems for the high-speed determination of physiologically important cationic or anionic electrolytes in body fluids. Example of flow injection determination of physiologically potassium in serum.  Several designs of low-volume potentiometric flow detectors have been reported. Flow-through potentiometric cell design
  • 25. Potentiometric microelectrodes are very suitable for in vivo real time clinical monitoring of blood electrolytes, intracellular studies, in situ environmental surveillance or industrial process control. For example Simon’s group described the utility of a system for on-line measurement of blood potassium ion concentration during an open-heart surgery. Miniaturized catheter-type ISE sensors such as implantable probe represent the preferred approach for routine clinical in vivo monitoring of blood electrolyte. Diamond's group developed an array of miniaturized chloride, sodium, and potassium ISEs for point-of-care analysis of sweat in connection to non- invasive diagnosis of cystic fibrosis. Miniaturized catheter-type ISE sensor
  • 26. Recent Advancement of Ion Selective Electrode (ISE)
  • 27. Agriculture and Fishery Pollution Control -Soil and fertilizer for -pH of acid rain, soil, surface water Nitrate, Ammonium, -Contamination of surface water Potassium to optimize the use of and ground water with ammonium fertilizer. and nitrate -Dissolved Oxygen and pH in ponds -Contamination of waste water with for Cyanide, Cadmium, Mercury and fish breeding. Copper Industrial Production Food and Quality Control -Salinity and pH of Boiler feed Worldwide -Nitrate and Nitrite in meat and application vegetables water -Chloride, Sodium, Nitrate and -Cyanide in plating baths Nitrite in -Process (specific ions) baby food. -Cadmium in fish Medical Diagnosis and Hygiene Control -Potassium in urine -Contamination in various ions.
  • 28. Advantages of Ion Selective Electrode (ISE) Technique  When compared to many other analytical techniques, Ion- Selective Electrodes are relatively inexpensive and simple to use and have an extremely wide range of applications and wide concentration range.  Under the most favourable conditions, when measuring ions in relatively dilute aqueous solutions and where interfering ions are not a problem, they can be used very rapidly and easily.  They are particularly useful in applications where only an order of magnitude concentration is required, or it is only necessary to know that a particular ion is below a certain concentration level.
  • 29. They are invaluable for the continuous monitoring of changes in concentration for example in potentiometric titrations or monitoring the uptake of nutrients, or the consumption of reagents.  They are particularly useful in biological/medical applications because they measure the activity of the ion directly, rather than the concentration.  ISEs are one of the few techniques which can measure both positive and negative ions.  They are unaffected by sample colour or turbidity.  ISEs can be used in aqueous solutions over a wide temperature range. Crystal membranes can operate in the range 0 C to 80 C and plastic membranes from 0 C to 50 C.
  • 30. Surface Plasmon Resonance (SPR)  Based on immunosensors for detection and monitoring of low-molecular- weight analytes of biomedical, food and environmental fields.  SPR is a surface sensitive optical technique for monitoring biomolecular interactions occurring in very close vicinity of a transducer (gold) surface, and that has given it a great potential for studying surface-confined affinity interactions without rinsing out unreacted or excess reactants in sample solutions.  It allows real-time study of the binding interactions between a biomolecule (antibody) immobilized on a transducer surface with its biospecific partner (analyte) in solution without the need for labeling the biomolecules by exploiting the interfacial refractive index changes associated with any affinity binding interaction.  In general, an SPR immunosensor is comprised of several important components: a light source, a detector, a transduction surface (usually gold-film), a prism, biomolecule (antibody or antigen) and a flow system.
  • 31. • The transduction surface is usually a thin gold-film (50–100 nm) on a glass slide optically coupled to a glass prism through a refractive index matching oil. In addition to gold, several metals can be used including silver, copper and aluminium. However, gold is highly preferred due to its chemical stability and free electron behaviour.
  • 32. Thank you for your attention