Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Redesign of Free Force Sensor Pinching Unit : GRASP
VA Rehabilitation Research & Development Center, Palo Alto
July-August 2005
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Project Team
IIT Kanpur VA Rehabilitation Research &
Development Center
Aakash Kumar
Ruchin Chaudhary
Niels Smaby
Mary Elise Johanson
Wendy Marie Murray
Eric Topp
Deborah Ellen Kenney
Stanford University
Maurice Leblanc
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Project Team
IIT Kanpur VA Rehabilitation Research &
Development Center
Aakash Kumar
Ruchin Chaudhary
Niels Smaby
Mary Elise Johanson
Wendy Marie Murray
Eric Topp
Deborah Ellen Kenney
Stanford University
Maurice Leblanc
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Project Team
IIT Kanpur VA Rehabilitation Research &
Development Center
Aakash Kumar
Ruchin Chaudhary
Niels Smaby
Mary Elise Johanson
Wendy Marie Murray
Eric Topp
Deborah Ellen Kenney
Stanford University
Maurice Leblanc
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Project Brief
System Objectives
System Schematics
System Analysis
Need Statement
Specifications & Constraints
Primary Product Requirement
Ergonomic Analysis
Analysis of Existing Design
References and Literature
Concept Generation
Discussions/Interactions
Physical mock-ups, Prototypes
Preliminary User Testing & Analysis
Detailing and validation
Final Prototype
Final Prototype Analysis
Discussion
Future Scope*
Project Methodology: Schedule
22nd June
29th June
1st July
4th-20th July
20th July
26th July
27th July- 11th
August
16th August
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Tetraplegics have restricted limb
movements
• The tendon transfer surgery for
hand can be helpful,it assist them
to apply forces through thumb and
index finger*
• Reconstructive hand surgery
enables the “key-pinch” in the
subjects
• Research is oriented towards
these pinch force measurements
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Medical Application
Schematics for surgery and functioning
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Tetraplegics have restricted limb
movements
• The tendon transfer surgery for
hand can be helpful,it assist them
to apply forces through thumb and
index finger*
• Reconstructive hand surgery
enables the “key-pinch” in the
subjects
• Research is oriented towards
these pinch force measurements
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Medical Application
Schematics for surgery and functioning
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Tetraplegics have restricted limb
movements
• The tendon transfer surgery for
hand can be helpful,it assist them
to apply forces through thumb and
index finger*
• Reconstructive hand surgery
enables the “key-pinch” in the
subjects
• Research is oriented towards
these pinch force measurements
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Medical Application
Schematics for surgery and functioning
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Tetraplegics have restricted limb
movements
• The tendon transfer surgery for
hand can be helpful,it assist them
to apply forces through thumb and
index finger*
• Reconstructive hand surgery
enables the “key-pinch” in the
subjects
• Research is oriented towards
these pinch force measurements
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Medical Application
Schematics for surgery and functioning
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Activities of daily living (ADLs)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Activities of daily living (ADLs)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Activities of daily living (ADLs)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Activities of daily living (ADLs)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Activities of daily living (ADLs)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Activities of daily living (ADLs)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The product is an assistive device to a study that aims to establish target
pinch forces for completing selected tasks that represent a range of
useful functional activities in daily life*.
• Opening and closing two types of zippers
• Inserting and removing a plug
• Inserting and removing a key
• Inserting and removing an ATM card
• Stabbing “food” with a fork
• Using a remote control button
*Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA;
Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD
System Objectives- Bio-Medical Research
Activities of daily living (ADLs)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
“The pinching unit or the GRASP is the clinical testing simulator of ADLs
(activities of daily livings) tasks with variable openings and lateral
adjustments for different individuals”
System Objectives- Product Application
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
System Schematics : Lateral pinch testing
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
“To re-design a comfortable* physical interface for a lateral
pinch testing medical equipment”
* Comfortable to both, the Subject and to the Clinician
Re-Design Need Statement
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Quantitative
1) Must be adjustable vertically from 14mm to 40mm
2) Should weigh less than 100gms
3) Must accommodate FF Sensor -Dimension Ø 40mm, Ht: 12.25mm/
Dimension Ø 17mm, Ht: 14.50mm
Qualitative
4) Must maintain a comfortable lateral pinch posture
5) Should be secured properly to the hand
6) Must have quick lateral adjustment
7) Must accommodate a thumb lateral angular calibration
8) Should interface with the micro scribe
Specifications and Constraints
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment: 14-40mm
b) Lateral pinch posture with hand securement
c) Quick lateral adjustment (proximal-distal adjustment)
d) Thumb orientation Calibration
e) Space orientation measurement
Keeping the comfort of the users (Subjects as well as Clinicians) into account
Primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment: 14-40mm
b) Lateral pinch posture with hand securement
c) Quick lateral adjustment (proximal-distal adjustment)
d) Thumb orientation Calibration
e) Space orientation measurement
Keeping the comfort of the users (Subjects as well as Clinicians) into account
Primary requirements
X
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment: 14-40mm
b) Lateral pinch posture with hand securement
c) Quick lateral adjustment (proximal-distal adjustment)
d) Thumb orientation Calibration
e) Space orientation measurement
Keeping the comfort of the users (Subjects as well as Clinicians) into account
Primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment: 14-40mm
b) Lateral pinch posture with hand securement
c) Quick lateral adjustment (proximal-distal adjustment)
d) Thumb orientation Calibration
e) Space orientation measurement
Keeping the comfort of the users (Subjects as well as Clinicians) into account
Primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment: 14-40mm
b) Lateral pinch posture with hand securement
c) Quick lateral adjustment (proximal-distal adjustment)
d) Thumb orientation Calibration
e) Space orientation measurement
Keeping the comfort of the users (Subjects as well as Clinicians) into account
Primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment: 14-40mm
b) Lateral pinch posture with hand securement
c) Quick lateral adjustment (proximal-distal adjustment)
d) Thumb orientation Calibration
e) Space orientation measurement
Keeping the comfort of the users (Subjects as well as Clinicians) into account
Primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• 5%ile and 95%ile fingers and thumb data (male/female)
• Reach of the thumb and fingers (5-95-M-F)
• Comfortable grips and postures*
• Restricted and unrestricted thumb-finger movements
*Strength Data for Design Safety, Product Safety and Testing Group, Institute for Occupational Ergonomics and Division of Manufacturing
Engineering and Operations Management,University of Nottingham, University Park, Nottingham NG7 2RD
Ergonomics Analysis
Hand Anthropometrics
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment
• Minimum 3mm required distance is not achievable (sensor limitation)
• Additional caps for height adjustment –Time consuming and
complex(Management of additional elements required)
• Cap’s vertical wall obstruct the thumb for reaching the Sensor
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment
• Minimum 3mm required distance is not achievable (sensor limitation)
• Additional caps for height adjustment –Time consuming and
complex(Management of additional elements required)
• Cap’s vertical wall obstruct the thumb for reaching the Sensor
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
a) Vertical adjustment
• Minimum 3mm required distance is not achievable (sensor limitation)
• Additional caps for height adjustment –Time consuming and
complex(Management of additional elements required)
• Cap’s vertical wall obstruct the thumb for reaching the Sensor
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
b) Lateral pinch posture with hand securement
• Strap forcefully affects the pinching direction
• Obstructs with the sensor as well as the hand at different positions
• Grip??? Is it required?
• Grip angle and girth might not cater to the entire population
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
b) Lateral pinch posture with hand securement
• Strap forcefully affects the pinching direction
• Obstructs with the sensor as well as the hand at different positions
• Grip??? Is it required?
• Grip angle and girth might not cater to the entire population
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
b) Lateral pinch posture with hand securement
• Strap forcefully affects the pinching direction
• Obstructs with the sensor as well as the hand at different positions
• Grip??? Is it required?
• Grip angle and girth might not cater to the entire population
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
b) Lateral pinch posture with hand securement
• Strap forcefully affects the pinching direction
• Obstructs with the sensor as well as the hand at different positions
• Grip??? Is it required?
• Grip angle and girth might not cater to the entire population
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
c) Quick lateral adjustment (proximal-distal adjustment)
• The current distance adjustment is 11mm, required is 20mm(5-95%ile data)
• Current adjustment requires screws and dismantling entire pinching
assembly- takes time
• Is it really required or can ergonomics can take care of it?
Existing Design: Analysis w.r.t. primary requirements
Proximal-Distal adjustment in existing design
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
c) Quick lateral adjustment (proximal-distal adjustment)
• The current distance adjustment is 11mm, required is 20mm(5-95%ile data)
• Current adjustment requires screws and dismantling entire pinching
assembly- takes time
• Is it really required or can ergonomics can take care of it?
Existing Design: Analysis w.r.t. primary requirements
Proximal-Distal adjustment in existing design & assembly system
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
c) Quick lateral adjustment (proximal-distal adjustment)
• The current distance adjustment is 11mm, required is 20mm(5-95%ile data)
• Current adjustment requires screws and dismantling entire pinching
assembly- takes time
• Is it really required or can ergonomics can take care of it?
Existing Design: Analysis w.r.t. primary requirements
Proximal-Distal adjustment in existing design & hand’s natural calibration
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
d) Thumb orientation measurement
• Transparent protractor printed on OHP sheet.
• Redesign needs to accommodate the calibration within the product
e) Space orientation measurement
• Design must mate with existing micro scribe for calibration
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
d) Thumb orientation measurement
• Transparent protractor printed on OHP sheet.
• Redesign needs to accommodate the calibration within the product
e) Space orientation measurement
• Design must mate with existing micro scribe for calibration
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
d) Thumb orientation measurement
• Transparent protractor printed on OHP sheet.
• Redesign needs to accommodate the calibration within the product
e) Space orientation measurement
• Design must mate with existing micro scribe for calibration
Existing Design: Analysis w.r.t. primary requirements
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 1 Modified Existing-1
• With Ergonomic Grip design
• Micro scribe attachment in grip
• To be strapped to the fingers
• With caps
• Existing modified-mates with the
current handle
• Handle can be rotated for better
fit
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 2 Modified Existing-2
• With Ergonomic Grip design
• Micro scribe attachment in grip
• To be strapped to the fingers
• With caps
• Existing modified-mates with the
current handle
• Thumb contours for maximum
reach
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 3 Telescopic tube grip less
• Without Grip design
• Micro scribe attachment in
front tube
• To be strapped to the fingers
• With telescopic shaft
• Close to the ADLs situation
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 4 Grip less strap design
• Without Grip design
• Micro scribe attachment in
platform
• To be strapped to the fingers
• With caps
• Close to the ADLs situation
• Natural calibration
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 5 Grip less snap design
• Without Grip design
• Micro scribe attachment in
platform sides
• To be tight-snapped to the
fingers
• With caps
• Close to the ADLs situation
• Natural calibration
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 6 Grip less ring design
• Without Grip design
• Micro scribe attachment in
platform-front
• To be worn on the fingers
• With caps
• Close to the ADLs situation
• Natural calibration
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 7 Thumb Grip design
• Without Grip design
• Micro scribe attachment in
platform-front
• To be strapped to the wrist
• With caps
• Close to the ADLs situation
• Natural calibration
• Additional stability around
thumb
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
With grip and without grip issue
Already a lot of patients have been tested “with grip” pre-operatively, thus a
change in the entire system would not be desirable right now. So the primary
design would focus on the ‘with grip design’. And the team can possibly explore
the other concepts for future scope.
Necessary improvements in the discussed concepts:
1) Fillets for avoiding thumb hitting the edges
2) Hand Grip adjustment
3) Tactile improvements for the grip surface
4) Strap section to act like a grip in itself(preferable)
5) Attach straps to the grip less and test it for functionality
Chosen Concepts for further detailing: 2, 4, 7
1st Discussions/Iterations/Clinician’s Interaction - 7/20/2005
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 2
• With Grip design
• Micro scribe attachment in grip
• To be strapped to the fingers
• Thumb profile contours at top
• With caps
• Existing modified-mates with the
current handle
Prototyping/Physical Mock-ups
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 4
• Grip-less design
• Micro scribe attachment
• Finger guide
• To be strapped to the fingers
• With Caps/Without caps
Prototyping/Physical Mock-ups
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Concept 7
• Grip-less design
• Micro scribe attachment in front
• To be strapped to the wrist
• Cushioning at the hand
interaction with device
• With caps
Prototyping/Physical Mock-ups
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Number of subjects-2
(1male,1female)
• Extreme cases-one close to
maximum and other minimum
• Mixed reactions to the prototypes
• Contour ergonomics felt
comfortable to the subjects in all the
cases better than the existing
• Strapping needs to be stabilize
and to be in soft Velcro (clinician’s
suggestion)
Preliminary User testing and analysis 07/26/2005-SCI Lab. VAPAHCS
User testing at SCI Lab
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Primary
requirements
Concept 4 Concept 7 Concept 2
Vertical adjustment 14-40mm (closest to
ADLs)
14-40mm (close to
ADLs)
14-40mm (with grip)
Lateral pinch posture
with hand securement
Unstable. Movement
with straps. Secured to
forefingers
Stable. Slight
movement with strap.
Secured at wrist and
thumb
Stable. No movement.
Secured to forefingers
and palm
Quick lateral adjustment
(proximal-distal
adjustment)
Not required Not required. Can be
incorporated
Required
Thumb orientation
measurement
On the platform chassis On the platform chassis On the platform chassis
Space orientation
measurement
Mate with micro-scribe
(in front)
Mate with micro-scribe
(in front)
Mate with micro-scribe
(in grip)
User’s Opinion
(subjects)
Fair Good Good
User’s Opinion
(Clinicians)
Good Fair Good
Detailing and Validation- Selection Criteria
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
The final prototype selected from the above comparative
analysis for further detailing is concept 2
Detailing and Validation- Selection Criteria
Evolution of concept-2 throughout the project
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Sketches to dimensioned drawings
• Drawings to AutoCAD data files for slices
• AutoCAD dxf files to Laser cutter
• SolidWorks model from AutoCAD data
• STL/IGES data exchange for CNC milling
(possible CAM application)
• Laser cutter slices assembly and finishing
• Velcro assembly and fittings /Alternate
details
Detailing and Validation- CAD/CAM
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype
Images showing the comfortable strapping and grip posture
Adjustability for lateral movement through caps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype – Working Drawings/ Platform
`
Platforms with High Visibility Thumb
calibration
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype – Working Drawings/ Caps
Different sized caps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype – Cost Specifications
• Raw material (wood-Balsa, Bass) ___ U.S.$ 12.50
• Glues/Adhesives ___ U.S.$ 2.75
• Laser cutter running cost ($2.5/Hr) ___ U.S.$ 7.50
• Velcro Straps ___ U.S.$ 3.00
Total Prototype Cost ___ U.S.$ 25.75
(INR 1,112)
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the thumb (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Final Prototype: Comparative Analysis with Existing Design
• Eliminated the rings-only caps required
• Two angles for lateral pinch posture
• Compact proximal-distal adjustment
• Mates with the current micro-scribe
• Integrated Thumb calibration provided
• Comfortable profile for the subject (fillets)
• Better securement to the hand
• Close to the correct position of Key-Pinch
• Less interchangeable parts for Clinicians
• Quick and easy adjustment of straps
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Better finish prototype with CNC milling or rapid prototyping
(assembly drawings ready)
• Minimum 3mm distance can be achieved by cantilever and 2nd lever
application (Conceptual sketches already discussed)
• Concept-7 can be developed further with CNC machined elastomeric
material for comfort fit (assembly drawings ready)
• Caps can be eliminated through telescopic assembly in the shaft,
which would provide quicker and compact assembly(Conceptual
sketches already discussed/optional)
• Robotic arm sensor can be calibrated with Gyro and Micro scribe to
be used as pinching unit
• Pinch pulling through friction can be tested for very small gap
openings
Future Scope
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Better finish prototype with CNC milling or rapid prototyping
(assembly drawings ready)
• Minimum 3mm distance can be achieved by cantilever and 2nd lever
application (Conceptual sketches already discussed)
• Concept-7 can be developed further with CNC machined elastomeric
material for comfort fit (assembly drawings ready)
• Caps can be eliminated through telescopic assembly in the shaft,
which would provide quicker and compact assembly(Conceptual
sketches already discussed/optional)
• Robotic arm sensor can be calibrated with Gyro and Micro scribe to
be used as pinching unit
• Pinch pulling through friction can be tested for very small gap
openings
Future Scope
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Better finish prototype with CNC milling or rapid prototyping
(assembly drawings ready)
• Minimum 3mm distance can be achieved by cantilever and 2nd lever
application (Conceptual sketches already discussed)
• Concept-7 can be developed further with CNC machined elastomeric
material for comfort fit (assembly drawings ready)
• Caps can be eliminated through telescopic assembly in the shaft,
which would provide quicker and compact assembly(Conceptual
sketches already discussed/optional)
• Robotic arm sensor can be calibrated with Gyro and Micro scribe to
be used as pinching unit
• Pinch pulling through friction can be tested for very small gap
openings
Future Scope
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Better finish prototype with CNC milling or rapid prototyping
(assembly drawings ready)
• Minimum 3mm distance can be achieved by cantilever and 2nd lever
application (Conceptual sketches already discussed)
• Concept-7 can be developed further with CNC machined elastomeric
material for comfort fit (assembly drawings ready)
• Caps can be eliminated through telescopic assembly in the shaft,
which would provide quicker and compact assembly(Conceptual
sketches already discussed/optional)
• Robotic arm sensor can be calibrated with Gyro and Micro scribe to
be used as pinching unit
• Pinch pulling through friction can be tested for very small gap
openings
Future Scope
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Better finish prototype with CNC milling or rapid prototyping
(assembly drawings ready)
• Minimum 3mm distance can be achieved by cantilever and 2nd lever
application (Conceptual sketches already discussed)
• Concept-7 can be developed further with CNC machined elastomeric
material for comfort fit (assembly drawings ready)
• Caps can be eliminated through telescopic assembly in the shaft,
which would provide quicker and compact assembly(Conceptual
sketches already discussed/optional)
• Robotic arm sensor can be calibrated with Gimbals and Micro scribe
to be used as pinching unit
• Pinch pulling through friction can be tested for very small gap
openings
Future Scope
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
• Better finish prototype with CNC milling or rapid prototyping
(assembly drawings ready)
• Minimum 3mm distance can be achieved by cantilever and 2nd lever
application (Conceptual sketches already discussed)
• Concept-7 can be developed further with CNC machined elastomeric
material for comfort fit (assembly drawings ready)
• Caps can be eliminated through telescopic assembly in the shaft,
which would provide quicker and compact assembly(Conceptual
sketches already discussed/optional)
• Robotic arm sensor can be calibrated with Gyro and Micro scribe to
be used as pinching unit
• Pinch pulling through friction can be tested for very small gap
openings
Future Scope
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
2nd Discussions/Iterations/Clinician’s Interaction - 8/16/2005
• Satisfied with the development of the final prototype
• Likes the idea, that improved design is in coherence with current system
• Soon the prototype would be developed in the final product (Manufacturing)
• Would like to observe the subject’s interaction with the final design, and would
notify Design team of results and necessary changes, if required
Pinching Unit Redesign – VA Rehabilitation Research & Development Center
Thank you !

Más contenido relacionado

Destacado

Seahorse nacho
Seahorse nachoSeahorse nacho
Seahorse nacho
Ainhoka
 
Прикумская фгос-начального-образования
Прикумская фгос-начального-образованияПрикумская фгос-начального-образования
Прикумская фгос-начального-образования
Anastasia Bulyga
 
Differences between baptism and dedication
Differences between baptism and dedicationDifferences between baptism and dedication
Differences between baptism and dedication
Lisa Rowland
 
Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...
Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...
Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...
Hector Del Castillo, CPM, CPMM
 

Destacado (20)

LETSCATCHUP UI 2016
LETSCATCHUP UI 2016LETSCATCHUP UI 2016
LETSCATCHUP UI 2016
 
Guinness - Engeneering Facts
Guinness - Engeneering FactsGuinness - Engeneering Facts
Guinness - Engeneering Facts
 
Daniel Blaufuks
Daniel BlaufuksDaniel Blaufuks
Daniel Blaufuks
 
Christofer Gilbert Artwork
Christofer Gilbert ArtworkChristofer Gilbert Artwork
Christofer Gilbert Artwork
 
Redondo Flower Fest
Redondo Flower FestRedondo Flower Fest
Redondo Flower Fest
 
Graphics branding 2016
Graphics branding 2016Graphics branding 2016
Graphics branding 2016
 
Howard Schatz
Howard SchatzHoward Schatz
Howard Schatz
 
Profile Dynamics Info
Profile Dynamics InfoProfile Dynamics Info
Profile Dynamics Info
 
Progragramacio curricular
Progragramacio curricularProgragramacio curricular
Progragramacio curricular
 
Pyatt Boardmark Investor Presentation Fund II
Pyatt Boardmark Investor Presentation Fund IIPyatt Boardmark Investor Presentation Fund II
Pyatt Boardmark Investor Presentation Fund II
 
Seahorse nacho
Seahorse nachoSeahorse nacho
Seahorse nacho
 
JJF2012とか?
JJF2012とか? JJF2012とか?
JJF2012とか?
 
Прикумская фгос-начального-образования
Прикумская фгос-начального-образованияПрикумская фгос-начального-образования
Прикумская фгос-начального-образования
 
Презентація класного керівника Гуцкалюк Л. І.
Презентація класного керівника Гуцкалюк Л. І.Презентація класного керівника Гуцкалюк Л. І.
Презентація класного керівника Гуцкалюк Л. І.
 
La navidad guadalupe
La navidad guadalupeLa navidad guadalupe
La navidad guadalupe
 
Towards an MDA Mechanism for RESTful Services Development
Towards an MDA Mechanism for RESTful Services DevelopmentTowards an MDA Mechanism for RESTful Services Development
Towards an MDA Mechanism for RESTful Services Development
 
Differences between baptism and dedication
Differences between baptism and dedicationDifferences between baptism and dedication
Differences between baptism and dedication
 
задачи
задачизадачи
задачи
 
Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...
Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...
Why Leading Innovation Is Essential For Product Managers - H. Del Castillo - ...
 
Hany ElMoghazy cv
Hany ElMoghazy cvHany ElMoghazy cv
Hany ElMoghazy cv
 

Similar a GRASP REDESIGN STANFORD

Introduction to the science of improving patient safety
Introduction to the science of improving patient safetyIntroduction to the science of improving patient safety
Introduction to the science of improving patient safety
ht3
 
public health _03302015 resized
public health _03302015 resizedpublic health _03302015 resized
public health _03302015 resized
Zakkoyya Lewis
 
Hinri Labs
Hinri LabsHinri Labs
Hinri Labs
jgoode73
 

Similar a GRASP REDESIGN STANFORD (20)

IoT, Big Data Healthcare Summit Western Canada - Wearables from Lab to Bedsid...
IoT, Big Data Healthcare Summit Western Canada - Wearables from Lab to Bedsid...IoT, Big Data Healthcare Summit Western Canada - Wearables from Lab to Bedsid...
IoT, Big Data Healthcare Summit Western Canada - Wearables from Lab to Bedsid...
 
Implanted Neural Prosthetics - an Introduction
Implanted Neural Prosthetics - an IntroductionImplanted Neural Prosthetics - an Introduction
Implanted Neural Prosthetics - an Introduction
 
Fitabase at 2016 Fitbit Captivate Summit
Fitabase at 2016 Fitbit Captivate SummitFitabase at 2016 Fitbit Captivate Summit
Fitabase at 2016 Fitbit Captivate Summit
 
RIWC_PARA_A029 Gait Analysis
RIWC_PARA_A029 Gait Analysis RIWC_PARA_A029 Gait Analysis
RIWC_PARA_A029 Gait Analysis
 
Health and Wealth Research Summary-05-15
Health and Wealth Research Summary-05-15Health and Wealth Research Summary-05-15
Health and Wealth Research Summary-05-15
 
Augmenting Healthcare by Supporting General Practitioners and Disclosing Hea...
 Augmenting Healthcare by Supporting General Practitioners and Disclosing Hea... Augmenting Healthcare by Supporting General Practitioners and Disclosing Hea...
Augmenting Healthcare by Supporting General Practitioners and Disclosing Hea...
 
Introduction to the science of improving patient safety
Introduction to the science of improving patient safetyIntroduction to the science of improving patient safety
Introduction to the science of improving patient safety
 
Evidence Based Practice- Implications
 Evidence Based Practice- Implications Evidence Based Practice- Implications
Evidence Based Practice- Implications
 
Title
TitleTitle
Title
 
Beyond Reporting: Monitoring and Evaluation as a Health Systems Strengthening...
Beyond Reporting: Monitoring and Evaluation as a Health Systems Strengthening...Beyond Reporting: Monitoring and Evaluation as a Health Systems Strengthening...
Beyond Reporting: Monitoring and Evaluation as a Health Systems Strengthening...
 
public health _03302015 resized
public health _03302015 resizedpublic health _03302015 resized
public health _03302015 resized
 
How to Conduct Medical Research
How to Conduct Medical ResearchHow to Conduct Medical Research
How to Conduct Medical Research
 
Hinri Labs
Hinri LabsHinri Labs
Hinri Labs
 
Healthcare Monitoring Using Wireless Sensor Networks
Healthcare Monitoring Using Wireless Sensor NetworksHealthcare Monitoring Using Wireless Sensor Networks
Healthcare Monitoring Using Wireless Sensor Networks
 
Volchenboum evolution summit_no_phi
Volchenboum evolution summit_no_phiVolchenboum evolution summit_no_phi
Volchenboum evolution summit_no_phi
 
IRB Approval Process 2016
IRB Approval Process 2016IRB Approval Process 2016
IRB Approval Process 2016
 
Psychologists and Quality Improvement 3.pdf
Psychologists and Quality Improvement 3.pdfPsychologists and Quality Improvement 3.pdf
Psychologists and Quality Improvement 3.pdf
 
Alzheimer’s & Parkinson’s Disease: Open Access
Alzheimer’s & Parkinson’s Disease: Open AccessAlzheimer’s & Parkinson’s Disease: Open Access
Alzheimer’s & Parkinson’s Disease: Open Access
 
Big Data and Practice-based Evidence: How EHR data is bringing the voice of n...
Big Data and Practice-based Evidence: How EHR data is bringing the voice of n...Big Data and Practice-based Evidence: How EHR data is bringing the voice of n...
Big Data and Practice-based Evidence: How EHR data is bringing the voice of n...
 
4. Determinants of health.pptx
4. Determinants of health.pptx4. Determinants of health.pptx
4. Determinants of health.pptx
 

Más de Ruchin Chaudhary (6)

Design suggestion presentation for central park bhilai
Design suggestion presentation for central park bhilaiDesign suggestion presentation for central park bhilai
Design suggestion presentation for central park bhilai
 
Letscatchup ui 2016
Letscatchup ui 2016Letscatchup ui 2016
Letscatchup ui 2016
 
Nokia eyeMitra
Nokia eyeMitraNokia eyeMitra
Nokia eyeMitra
 
NIRAMAYA CSR PRESENTATION 290614.PDF
NIRAMAYA CSR PRESENTATION 290614.PDFNIRAMAYA CSR PRESENTATION 290614.PDF
NIRAMAYA CSR PRESENTATION 290614.PDF
 
RUCHIN IITK THESIS
RUCHIN IITK THESISRUCHIN IITK THESIS
RUCHIN IITK THESIS
 
DESIGN PROFILE 2016
DESIGN PROFILE 2016DESIGN PROFILE 2016
DESIGN PROFILE 2016
 

GRASP REDESIGN STANFORD

  • 1. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Redesign of Free Force Sensor Pinching Unit : GRASP VA Rehabilitation Research & Development Center, Palo Alto July-August 2005
  • 2. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Project Team IIT Kanpur VA Rehabilitation Research & Development Center Aakash Kumar Ruchin Chaudhary Niels Smaby Mary Elise Johanson Wendy Marie Murray Eric Topp Deborah Ellen Kenney Stanford University Maurice Leblanc
  • 3. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Project Team IIT Kanpur VA Rehabilitation Research & Development Center Aakash Kumar Ruchin Chaudhary Niels Smaby Mary Elise Johanson Wendy Marie Murray Eric Topp Deborah Ellen Kenney Stanford University Maurice Leblanc
  • 4. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Project Team IIT Kanpur VA Rehabilitation Research & Development Center Aakash Kumar Ruchin Chaudhary Niels Smaby Mary Elise Johanson Wendy Marie Murray Eric Topp Deborah Ellen Kenney Stanford University Maurice Leblanc
  • 5. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Project Brief System Objectives System Schematics System Analysis Need Statement Specifications & Constraints Primary Product Requirement Ergonomic Analysis Analysis of Existing Design References and Literature Concept Generation Discussions/Interactions Physical mock-ups, Prototypes Preliminary User Testing & Analysis Detailing and validation Final Prototype Final Prototype Analysis Discussion Future Scope* Project Methodology: Schedule 22nd June 29th June 1st July 4th-20th July 20th July 26th July 27th July- 11th August 16th August
  • 6. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Tetraplegics have restricted limb movements • The tendon transfer surgery for hand can be helpful,it assist them to apply forces through thumb and index finger* • Reconstructive hand surgery enables the “key-pinch” in the subjects • Research is oriented towards these pinch force measurements *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Medical Application Schematics for surgery and functioning
  • 7. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Tetraplegics have restricted limb movements • The tendon transfer surgery for hand can be helpful,it assist them to apply forces through thumb and index finger* • Reconstructive hand surgery enables the “key-pinch” in the subjects • Research is oriented towards these pinch force measurements *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Medical Application Schematics for surgery and functioning
  • 8. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Tetraplegics have restricted limb movements • The tendon transfer surgery for hand can be helpful,it assist them to apply forces through thumb and index finger* • Reconstructive hand surgery enables the “key-pinch” in the subjects • Research is oriented towards these pinch force measurements *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Medical Application Schematics for surgery and functioning
  • 9. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Tetraplegics have restricted limb movements • The tendon transfer surgery for hand can be helpful,it assist them to apply forces through thumb and index finger* • Reconstructive hand surgery enables the “key-pinch” in the subjects • Research is oriented towards these pinch force measurements *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Medical Application Schematics for surgery and functioning
  • 10. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research
  • 11. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research Activities of daily living (ADLs)
  • 12. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research Activities of daily living (ADLs)
  • 13. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research Activities of daily living (ADLs)
  • 14. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research Activities of daily living (ADLs)
  • 15. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research Activities of daily living (ADLs)
  • 16. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research Activities of daily living (ADLs)
  • 17. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The product is an assistive device to a study that aims to establish target pinch forces for completing selected tasks that represent a range of useful functional activities in daily life*. • Opening and closing two types of zippers • Inserting and removing a plug • Inserting and removing a key • Inserting and removing an ATM card • Stabbing “food” with a fork • Using a remote control button *Journal of Rehabilitation Research and Development,Mar-Apr 2004 , Niels Smaby, MS; Mary Elise Johanson, MS, PT; Brian Baker, BA; Deborah Ellen Kenney, MS, OTR;Wendy Marie Murray, MS, PhD; Vincent Rod Hentz, MD System Objectives- Bio-Medical Research Activities of daily living (ADLs)
  • 18. Pinching Unit Redesign – VA Rehabilitation Research & Development Center “The pinching unit or the GRASP is the clinical testing simulator of ADLs (activities of daily livings) tasks with variable openings and lateral adjustments for different individuals” System Objectives- Product Application
  • 19. Pinching Unit Redesign – VA Rehabilitation Research & Development Center System Schematics : Lateral pinch testing
  • 20. Pinching Unit Redesign – VA Rehabilitation Research & Development Center “To re-design a comfortable* physical interface for a lateral pinch testing medical equipment” * Comfortable to both, the Subject and to the Clinician Re-Design Need Statement
  • 21. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Quantitative 1) Must be adjustable vertically from 14mm to 40mm 2) Should weigh less than 100gms 3) Must accommodate FF Sensor -Dimension Ø 40mm, Ht: 12.25mm/ Dimension Ø 17mm, Ht: 14.50mm Qualitative 4) Must maintain a comfortable lateral pinch posture 5) Should be secured properly to the hand 6) Must have quick lateral adjustment 7) Must accommodate a thumb lateral angular calibration 8) Should interface with the micro scribe Specifications and Constraints
  • 22. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment: 14-40mm b) Lateral pinch posture with hand securement c) Quick lateral adjustment (proximal-distal adjustment) d) Thumb orientation Calibration e) Space orientation measurement Keeping the comfort of the users (Subjects as well as Clinicians) into account Primary requirements
  • 23. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment: 14-40mm b) Lateral pinch posture with hand securement c) Quick lateral adjustment (proximal-distal adjustment) d) Thumb orientation Calibration e) Space orientation measurement Keeping the comfort of the users (Subjects as well as Clinicians) into account Primary requirements X
  • 24. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment: 14-40mm b) Lateral pinch posture with hand securement c) Quick lateral adjustment (proximal-distal adjustment) d) Thumb orientation Calibration e) Space orientation measurement Keeping the comfort of the users (Subjects as well as Clinicians) into account Primary requirements
  • 25. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment: 14-40mm b) Lateral pinch posture with hand securement c) Quick lateral adjustment (proximal-distal adjustment) d) Thumb orientation Calibration e) Space orientation measurement Keeping the comfort of the users (Subjects as well as Clinicians) into account Primary requirements
  • 26. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment: 14-40mm b) Lateral pinch posture with hand securement c) Quick lateral adjustment (proximal-distal adjustment) d) Thumb orientation Calibration e) Space orientation measurement Keeping the comfort of the users (Subjects as well as Clinicians) into account Primary requirements
  • 27. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment: 14-40mm b) Lateral pinch posture with hand securement c) Quick lateral adjustment (proximal-distal adjustment) d) Thumb orientation Calibration e) Space orientation measurement Keeping the comfort of the users (Subjects as well as Clinicians) into account Primary requirements
  • 28. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • 5%ile and 95%ile fingers and thumb data (male/female) • Reach of the thumb and fingers (5-95-M-F) • Comfortable grips and postures* • Restricted and unrestricted thumb-finger movements *Strength Data for Design Safety, Product Safety and Testing Group, Institute for Occupational Ergonomics and Division of Manufacturing Engineering and Operations Management,University of Nottingham, University Park, Nottingham NG7 2RD Ergonomics Analysis Hand Anthropometrics
  • 29. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment • Minimum 3mm required distance is not achievable (sensor limitation) • Additional caps for height adjustment –Time consuming and complex(Management of additional elements required) • Cap’s vertical wall obstruct the thumb for reaching the Sensor Existing Design: Analysis w.r.t. primary requirements
  • 30. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment • Minimum 3mm required distance is not achievable (sensor limitation) • Additional caps for height adjustment –Time consuming and complex(Management of additional elements required) • Cap’s vertical wall obstruct the thumb for reaching the Sensor Existing Design: Analysis w.r.t. primary requirements
  • 31. Pinching Unit Redesign – VA Rehabilitation Research & Development Center a) Vertical adjustment • Minimum 3mm required distance is not achievable (sensor limitation) • Additional caps for height adjustment –Time consuming and complex(Management of additional elements required) • Cap’s vertical wall obstruct the thumb for reaching the Sensor Existing Design: Analysis w.r.t. primary requirements
  • 32. Pinching Unit Redesign – VA Rehabilitation Research & Development Center b) Lateral pinch posture with hand securement • Strap forcefully affects the pinching direction • Obstructs with the sensor as well as the hand at different positions • Grip??? Is it required? • Grip angle and girth might not cater to the entire population Existing Design: Analysis w.r.t. primary requirements
  • 33. Pinching Unit Redesign – VA Rehabilitation Research & Development Center b) Lateral pinch posture with hand securement • Strap forcefully affects the pinching direction • Obstructs with the sensor as well as the hand at different positions • Grip??? Is it required? • Grip angle and girth might not cater to the entire population Existing Design: Analysis w.r.t. primary requirements
  • 34. Pinching Unit Redesign – VA Rehabilitation Research & Development Center b) Lateral pinch posture with hand securement • Strap forcefully affects the pinching direction • Obstructs with the sensor as well as the hand at different positions • Grip??? Is it required? • Grip angle and girth might not cater to the entire population Existing Design: Analysis w.r.t. primary requirements
  • 35. Pinching Unit Redesign – VA Rehabilitation Research & Development Center b) Lateral pinch posture with hand securement • Strap forcefully affects the pinching direction • Obstructs with the sensor as well as the hand at different positions • Grip??? Is it required? • Grip angle and girth might not cater to the entire population Existing Design: Analysis w.r.t. primary requirements
  • 36. Pinching Unit Redesign – VA Rehabilitation Research & Development Center c) Quick lateral adjustment (proximal-distal adjustment) • The current distance adjustment is 11mm, required is 20mm(5-95%ile data) • Current adjustment requires screws and dismantling entire pinching assembly- takes time • Is it really required or can ergonomics can take care of it? Existing Design: Analysis w.r.t. primary requirements Proximal-Distal adjustment in existing design
  • 37. Pinching Unit Redesign – VA Rehabilitation Research & Development Center c) Quick lateral adjustment (proximal-distal adjustment) • The current distance adjustment is 11mm, required is 20mm(5-95%ile data) • Current adjustment requires screws and dismantling entire pinching assembly- takes time • Is it really required or can ergonomics can take care of it? Existing Design: Analysis w.r.t. primary requirements Proximal-Distal adjustment in existing design & assembly system
  • 38. Pinching Unit Redesign – VA Rehabilitation Research & Development Center c) Quick lateral adjustment (proximal-distal adjustment) • The current distance adjustment is 11mm, required is 20mm(5-95%ile data) • Current adjustment requires screws and dismantling entire pinching assembly- takes time • Is it really required or can ergonomics can take care of it? Existing Design: Analysis w.r.t. primary requirements Proximal-Distal adjustment in existing design & hand’s natural calibration
  • 39. Pinching Unit Redesign – VA Rehabilitation Research & Development Center d) Thumb orientation measurement • Transparent protractor printed on OHP sheet. • Redesign needs to accommodate the calibration within the product e) Space orientation measurement • Design must mate with existing micro scribe for calibration Existing Design: Analysis w.r.t. primary requirements
  • 40. Pinching Unit Redesign – VA Rehabilitation Research & Development Center d) Thumb orientation measurement • Transparent protractor printed on OHP sheet. • Redesign needs to accommodate the calibration within the product e) Space orientation measurement • Design must mate with existing micro scribe for calibration Existing Design: Analysis w.r.t. primary requirements
  • 41. Pinching Unit Redesign – VA Rehabilitation Research & Development Center d) Thumb orientation measurement • Transparent protractor printed on OHP sheet. • Redesign needs to accommodate the calibration within the product e) Space orientation measurement • Design must mate with existing micro scribe for calibration Existing Design: Analysis w.r.t. primary requirements
  • 42. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 1 Modified Existing-1 • With Ergonomic Grip design • Micro scribe attachment in grip • To be strapped to the fingers • With caps • Existing modified-mates with the current handle • Handle can be rotated for better fit
  • 43. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 2 Modified Existing-2 • With Ergonomic Grip design • Micro scribe attachment in grip • To be strapped to the fingers • With caps • Existing modified-mates with the current handle • Thumb contours for maximum reach
  • 44. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 3 Telescopic tube grip less • Without Grip design • Micro scribe attachment in front tube • To be strapped to the fingers • With telescopic shaft • Close to the ADLs situation
  • 45. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 4 Grip less strap design • Without Grip design • Micro scribe attachment in platform • To be strapped to the fingers • With caps • Close to the ADLs situation • Natural calibration
  • 46. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 5 Grip less snap design • Without Grip design • Micro scribe attachment in platform sides • To be tight-snapped to the fingers • With caps • Close to the ADLs situation • Natural calibration
  • 47. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 6 Grip less ring design • Without Grip design • Micro scribe attachment in platform-front • To be worn on the fingers • With caps • Close to the ADLs situation • Natural calibration
  • 48. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 7 Thumb Grip design • Without Grip design • Micro scribe attachment in platform-front • To be strapped to the wrist • With caps • Close to the ADLs situation • Natural calibration • Additional stability around thumb
  • 49. Pinching Unit Redesign – VA Rehabilitation Research & Development Center With grip and without grip issue Already a lot of patients have been tested “with grip” pre-operatively, thus a change in the entire system would not be desirable right now. So the primary design would focus on the ‘with grip design’. And the team can possibly explore the other concepts for future scope. Necessary improvements in the discussed concepts: 1) Fillets for avoiding thumb hitting the edges 2) Hand Grip adjustment 3) Tactile improvements for the grip surface 4) Strap section to act like a grip in itself(preferable) 5) Attach straps to the grip less and test it for functionality Chosen Concepts for further detailing: 2, 4, 7 1st Discussions/Iterations/Clinician’s Interaction - 7/20/2005
  • 50. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 2 • With Grip design • Micro scribe attachment in grip • To be strapped to the fingers • Thumb profile contours at top • With caps • Existing modified-mates with the current handle Prototyping/Physical Mock-ups
  • 51. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 4 • Grip-less design • Micro scribe attachment • Finger guide • To be strapped to the fingers • With Caps/Without caps Prototyping/Physical Mock-ups
  • 52. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Concept 7 • Grip-less design • Micro scribe attachment in front • To be strapped to the wrist • Cushioning at the hand interaction with device • With caps Prototyping/Physical Mock-ups
  • 53. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Number of subjects-2 (1male,1female) • Extreme cases-one close to maximum and other minimum • Mixed reactions to the prototypes • Contour ergonomics felt comfortable to the subjects in all the cases better than the existing • Strapping needs to be stabilize and to be in soft Velcro (clinician’s suggestion) Preliminary User testing and analysis 07/26/2005-SCI Lab. VAPAHCS User testing at SCI Lab
  • 54. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Primary requirements Concept 4 Concept 7 Concept 2 Vertical adjustment 14-40mm (closest to ADLs) 14-40mm (close to ADLs) 14-40mm (with grip) Lateral pinch posture with hand securement Unstable. Movement with straps. Secured to forefingers Stable. Slight movement with strap. Secured at wrist and thumb Stable. No movement. Secured to forefingers and palm Quick lateral adjustment (proximal-distal adjustment) Not required Not required. Can be incorporated Required Thumb orientation measurement On the platform chassis On the platform chassis On the platform chassis Space orientation measurement Mate with micro-scribe (in front) Mate with micro-scribe (in front) Mate with micro-scribe (in grip) User’s Opinion (subjects) Fair Good Good User’s Opinion (Clinicians) Good Fair Good Detailing and Validation- Selection Criteria
  • 55. Pinching Unit Redesign – VA Rehabilitation Research & Development Center The final prototype selected from the above comparative analysis for further detailing is concept 2 Detailing and Validation- Selection Criteria Evolution of concept-2 throughout the project
  • 56. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Sketches to dimensioned drawings • Drawings to AutoCAD data files for slices • AutoCAD dxf files to Laser cutter • SolidWorks model from AutoCAD data • STL/IGES data exchange for CNC milling (possible CAM application) • Laser cutter slices assembly and finishing • Velcro assembly and fittings /Alternate details Detailing and Validation- CAD/CAM
  • 57. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype Images showing the comfortable strapping and grip posture Adjustability for lateral movement through caps
  • 58. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype – Working Drawings/ Platform ` Platforms with High Visibility Thumb calibration
  • 59. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype – Working Drawings/ Caps Different sized caps
  • 60. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype – Cost Specifications • Raw material (wood-Balsa, Bass) ___ U.S.$ 12.50 • Glues/Adhesives ___ U.S.$ 2.75 • Laser cutter running cost ($2.5/Hr) ___ U.S.$ 7.50 • Velcro Straps ___ U.S.$ 3.00 Total Prototype Cost ___ U.S.$ 25.75 (INR 1,112)
  • 61. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 62. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 63. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 64. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 65. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 66. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the thumb (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 67. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 68. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 69. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 70. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 71. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Final Prototype: Comparative Analysis with Existing Design • Eliminated the rings-only caps required • Two angles for lateral pinch posture • Compact proximal-distal adjustment • Mates with the current micro-scribe • Integrated Thumb calibration provided • Comfortable profile for the subject (fillets) • Better securement to the hand • Close to the correct position of Key-Pinch • Less interchangeable parts for Clinicians • Quick and easy adjustment of straps
  • 72. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Better finish prototype with CNC milling or rapid prototyping (assembly drawings ready) • Minimum 3mm distance can be achieved by cantilever and 2nd lever application (Conceptual sketches already discussed) • Concept-7 can be developed further with CNC machined elastomeric material for comfort fit (assembly drawings ready) • Caps can be eliminated through telescopic assembly in the shaft, which would provide quicker and compact assembly(Conceptual sketches already discussed/optional) • Robotic arm sensor can be calibrated with Gyro and Micro scribe to be used as pinching unit • Pinch pulling through friction can be tested for very small gap openings Future Scope
  • 73. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Better finish prototype with CNC milling or rapid prototyping (assembly drawings ready) • Minimum 3mm distance can be achieved by cantilever and 2nd lever application (Conceptual sketches already discussed) • Concept-7 can be developed further with CNC machined elastomeric material for comfort fit (assembly drawings ready) • Caps can be eliminated through telescopic assembly in the shaft, which would provide quicker and compact assembly(Conceptual sketches already discussed/optional) • Robotic arm sensor can be calibrated with Gyro and Micro scribe to be used as pinching unit • Pinch pulling through friction can be tested for very small gap openings Future Scope
  • 74. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Better finish prototype with CNC milling or rapid prototyping (assembly drawings ready) • Minimum 3mm distance can be achieved by cantilever and 2nd lever application (Conceptual sketches already discussed) • Concept-7 can be developed further with CNC machined elastomeric material for comfort fit (assembly drawings ready) • Caps can be eliminated through telescopic assembly in the shaft, which would provide quicker and compact assembly(Conceptual sketches already discussed/optional) • Robotic arm sensor can be calibrated with Gyro and Micro scribe to be used as pinching unit • Pinch pulling through friction can be tested for very small gap openings Future Scope
  • 75. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Better finish prototype with CNC milling or rapid prototyping (assembly drawings ready) • Minimum 3mm distance can be achieved by cantilever and 2nd lever application (Conceptual sketches already discussed) • Concept-7 can be developed further with CNC machined elastomeric material for comfort fit (assembly drawings ready) • Caps can be eliminated through telescopic assembly in the shaft, which would provide quicker and compact assembly(Conceptual sketches already discussed/optional) • Robotic arm sensor can be calibrated with Gyro and Micro scribe to be used as pinching unit • Pinch pulling through friction can be tested for very small gap openings Future Scope
  • 76. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Better finish prototype with CNC milling or rapid prototyping (assembly drawings ready) • Minimum 3mm distance can be achieved by cantilever and 2nd lever application (Conceptual sketches already discussed) • Concept-7 can be developed further with CNC machined elastomeric material for comfort fit (assembly drawings ready) • Caps can be eliminated through telescopic assembly in the shaft, which would provide quicker and compact assembly(Conceptual sketches already discussed/optional) • Robotic arm sensor can be calibrated with Gimbals and Micro scribe to be used as pinching unit • Pinch pulling through friction can be tested for very small gap openings Future Scope
  • 77. Pinching Unit Redesign – VA Rehabilitation Research & Development Center • Better finish prototype with CNC milling or rapid prototyping (assembly drawings ready) • Minimum 3mm distance can be achieved by cantilever and 2nd lever application (Conceptual sketches already discussed) • Concept-7 can be developed further with CNC machined elastomeric material for comfort fit (assembly drawings ready) • Caps can be eliminated through telescopic assembly in the shaft, which would provide quicker and compact assembly(Conceptual sketches already discussed/optional) • Robotic arm sensor can be calibrated with Gyro and Micro scribe to be used as pinching unit • Pinch pulling through friction can be tested for very small gap openings Future Scope
  • 78. Pinching Unit Redesign – VA Rehabilitation Research & Development Center 2nd Discussions/Iterations/Clinician’s Interaction - 8/16/2005 • Satisfied with the development of the final prototype • Likes the idea, that improved design is in coherence with current system • Soon the prototype would be developed in the final product (Manufacturing) • Would like to observe the subject’s interaction with the final design, and would notify Design team of results and necessary changes, if required
  • 79. Pinching Unit Redesign – VA Rehabilitation Research & Development Center Thank you !