Power Protection for Digital Medical
Imaging and Diagnostic Equipment
Revision 2
by Viswas Purani
White Paper 86
by Schneider Electric White Papers are now part of the Schneider Electric
white paper library produced by Schneider Electric’s Data Center Science Center
DCSC@Schneider-Electric.com
Medical imaging and diagnostic equipment (MIDE) is
increasingly being networked to Picture Archiving and
Communications Systems (PACS), Radiology Infor-
mation Systems (RIS), Hospital Information Systems
(HIS), and getting connected to the hospital intranet as
well as the Internet. Failing to implement the neces-
sary physical infrastructure can result in unexpected
downtime, and safety and compliance issues, which
translates into lost revenue and exposure to expensive
litigations, negatively affecting the bottom line. This
paper explains how to plan for physical infrastructure
when deploying medical imaging and diagnostic
equipment, with emphasis on power and cooling.
Executive summary
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 2
The proliferation of information technology (IT) and other high technologies into medical
imaging and diagnostic equipment over the years has resulted in the evolution of powerful
new devices in the field of diagnostics and interventional radiology. The information generat-
ed and carried by these images is crucial to the treatment in cardiology, neurology, oncology,
gynecology & obstetrics, orthopedics, surgery, and pulmonary medicine. These new
developments have helped in the early detection and treatment of diseases and significantly
improved patient care. A typical medical imaging and diagnostic equipment network is
illustrated in Figure 1. This equipment can be broken down into five broad sub-categories:
1. Modalities that capture or generate the images
2. Picture Archiving and Communications Systems (PACS) that store the generated
images and make them available to the physicians and nurses for diagnosis and
treatment
3. Radiology Information Systems (RIS) and Hospital Information Systems (HIS) that
monitor and manage the work flow of radiology departments and entire hospitals, all
the way from the patient check-in, to scheduling and billing, to generating electronic
medical records and management reporting
4. Computed Radiography (CR) that helps convert films to digital images on cassettes or
Digital Radiography (DR) that provides cassette-less digital images
5. Laser printers that print films when required and other peripherals
CT, MRI, etc.
RIS
Viewing
Workstations
Modality
HIS
Consulting Cardiologist
Consulting Physician
Internet VPN
Storage - PACS
Modalities are high-tech medical imaging systems including Computed Tomography (CT),
Magnetic Resonance Imaging (MRI), Positron Emission Tomography (PET), Ultrasound (US),
and Electro-Cardiogram (ECG). They get connected to a PACS and to RIS/HIS through
Local Area Networks (LANs), and/or Wireless Local Area Networks (WLANs), or Wide Area
Networks (WANs). PACS may have their own Storage Area Networks (SANs) or Network
Attached Storage (NAS) and the RIS and HIS may be made of several clusters of servers and
a number of workstations distributed in different departments of the hospitals. Because of
their numerous benefits combined with the enormous pressure on the hospitals to improve
quality of care, reduce errors, comply with federal regulations like HIPAA, and simultaneously
cut cost, adoption of all these technologies is inevitable, converting the traditional hospital
into a “digital enterprise”.
Introduction
Figure 1
Typical MIDE network
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 3
The backbone of this new digital hospital is a network made up of different modalities, PACS
and RIS/HIS, CR/DR, printers, and peripherals. This highly complex network and its compo-
nents has to comply with relevant standards like Digital Imaging and Communications in
Medicine (DICOM), Health Language Seven (HL7), Underwriters Laboratory (UL), Federal
Communications Commission (FCC), National Electrical Code (NEC), and other applicable
local and national codes such as BS7671:2001 (U.K.), NFC15-100 (France), and VDE
(Germany), as well as international such as CEI IEC 60364. This imposes a huge challenge
to the IT and the facilities manager to provide the right physical infrastructure. This physical
infrastructure has to be reliable, scalable, highly available, and manageable. It consists of:
1. Power systems such as uninterruptible power supplies (UPSs), power distribution
units (PDUs), isolation transformers, and generators to provide uninterrupted, condi-
tioned, clean power to the critical loads
2. Precision cooling systems that provide optimal environment by regulating temperature
and humidity
3. Racks that house the critical network equipment like servers, switches, routers, and
gateways
4. Physical security and fire protection systems
5. Cabling to interconnect equipment
6. Management systems to monitor and manage these systems, locally as well as re-
motely to ensure their satisfactory operation 7x24x365
7. Services to design, deliver, install, commission, operate and maintain these systems
Special attention should be given to the hospital wiring closets which allow networking of the
modalities to PACS and RIS/HIS as well as other workstations and peripherals within the
hospital premises. It is these backbone closets that support this complex hospital network
carrying critical data, voice, and video, keeping the network up and running.
The hospital power system is a large complex electrical system consisting of high voltage
transformers, automatic transfer switches (ATS), generators, isolation transformers, PDUs,
etc. This power system feeds a variety of electrical loads including lighting, heating, ventila-
tion air-conditioning (HVAC) systems, elevators, escalators, large pumps, fans, motors and
more. The random nature of these loads (turning on and off randomly) creates an unstable
power environment (i.e. sags and surges) that more sensitive imaging and diagnostic
equipment and other IT networks that support them, must endure. Healthcare companies
and hospitals faced with these challenges should engage partners like Schneider Electric
with engineering expertise to perform complete physical infrastructure assessments that
identify weaknesses and suggest corrective actions. This paper discusses and reviews the
challenges imposed on physical infrastructure while deploying medical imaging and diagnos-
tic equipment, with a focus on power and cooling.
Depending on the patient’s ailment, a physician can use different modalities for diagnosis and
treatment (i.e. X-Ray or CT for orthopedics, ECG or MRI for cardiology or Ultrasound for
obstetrics). These modalities can be broadly classified into two categories: portable and
stationary. Portables can be further classified into hand held (i.e. blood glucometer) and
trolley or cart-mounted (i.e. ultrasound) while stationary devices can be further classified into
desk-mounted (i.e. blood, urine analysis equipment) or floor-mounted (i.e. CT, MRI). Figure
2 illustrates a cart-mounted ultrasound and Figure 3 illustrates a floor-mounted MRI machine.
The floor-mounted, desk-mounted, and cart-mounted devices need the most physical
infrastructure planning.
Modalities
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 4
Environment
Modalities are generally used in an indoor office environment. The cart-mounted and desk-
mounted modalities generally use 120 VAC, 208 VAC or 230 VAC single phase power less
than 5 kVA1
. The floor mounted devices typically require 208 VAC, 400 VAC or 480 VAC
three-phase power, ranging from 20 kVA to 300 kVA or more. They require a lot of space
and often times have their own separate room within the main hospital building or adjacent to
it. They may be cooled with the building’s comfort air-conditioning system or they may have
a precision computer room air-conditioning (CRAC) system, which more tightly controls
temperature and humidity in the environment. Figure 4 illustrates a typical MRI facility with
floor mounted equipment.
Challenges
Modalities need to be protected from power anomalies that cause hardware failures such as
blown power supplies or printed circuit boards (PCBs) as well as from system software
crashes. Physical space is a major constraint for large modalities like CTs and MRIs, more
so in big hospitals in urban areas, as they have no room to expand. These modalities
consume a lot of power, so heat dissipation is a major challenge to the building cooling
system. Often times, comfort cooling is not sufficient and precision cooling is required. One
of the most critical requirements is to provide electrical isolation from the electrical utility input
to protect the patient and the technicians from any shock hazards. Compliance to local,
state, national, and international codes is of paramount importance (i.e. National Electrical
Codes (NEC) in the United States).
1
See White Paper 15, Watts and Volt-Amps: Powerful Confusion for information on kilo volt amps
Figure 2 (left)
Cart mounted ultrasound
Figure 3 (right)
Floor mounted MRI
Figure 4
Typical CT or MRI facility
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 5
Best practices
• Since the hospital power grid is electrically “noisy and dirty” with a lot of electrical
surges and sags, it is a good practice to provide UPS protection to all sensitive, expen-
sive electronics systems, LCD displays, workstations, printers, and peripherals. The
UPS system protects the hardware, avoids unwarranted system crashes while tests are
in progress, prevents loss of patient data files, and provides safe, reliable radiology ex-
aminations.
• UPS systems used in the hospital should meet the following stringent standards:
o UL1778 – American Standard for UPS
o CSA22.2 No. 107.1 – Canadian Standard of UPS
o FCC Part 15 Class A – American Standard for Electromagnetic Radiation
o ANSI C62.41- American Standard for surge withstand capability
o IEC60950 – International Standard for UPS Systems
o EN50091-1- European Standard for Electromagnetic compatibility
Depending on their usage, cart-mounted and desk-mounted devices may require the
UPS to comply with the international standard IEC60601-1, “Standard for Medical Elec-
trical Equipment, Part 1: General Requirements for Safety” for patient vicinity applica-
tions. The U.S. equivalent is UL 60601-1.
• For large floor-mounted modalities, a large UPS (50-300kVA typical) should be in-
stalled to protect the entire room. Isolation transformers and the appropriate circuit
breakers should be used to limit the leakage currents and electrical shock hazards for
such modalities. These isolation transformers are sometimes built into the UPS or may
be optional outside of the UPS box. All equipment should be installed in compliance
with relevant codes like NEC, NFPA 70, NFPA99 (for U.S), and all other applicable lo-
cal and national codes such as BS7671:2001 (U.K.), NFC15-100 (France), and VDE
(Germany), as well as international such as CEI IEC 60364. Understanding and inter-
preting the codes can, in some unique cases, become very controversial and the final
say should be from the authorities having jurisdiction. In instances where budget con-
straints prohibit investment in a large facility level UPS, a smaller UPS (5-10kVA) dedi-
cated to the sensitive electronics and computer system of the CT, MRI, and PET should
be considered. In addition, UPS protection should be provided to all their viewing sta-
tions & workstations.
• Sizing of the UPS for many devices like CTs and MRIs can be challenging since they
draw very high amounts of inrush current. Ample precautions should be taken while
sizing their power systems (including UPS, generators, transformers, and switchgear).
Their normal power consumption as well as the inrush current ratings are available
from their manufacturers. Allow enough margins for miscellaneous loads and future
growth. Many companies like Schneider Electric have dedicated systems engineers,
power protection specialists, etc. who can help evaluate the right solution for every
unique customer situation.
• Adequate cooling and air flow should be provided for all modalities that have sensitive
electronics dissipating heat. For most of the cart and desk-mounted modalities, build-
ing HVAC should be sufficient. However for large floor mounted modalities like CTs,
MRIs, or PETs supplemental cooling may be required. Precision cooling is preferred
as it can provide temperature and humidity control in the CT/MRI room.
• All of the networked modalities and their physical infrastructure should be monitored
and managed (i.e. environmental conditions of the radiology room, UPS battery life,
runtime and capacity, and generator fuel) so that anomalies can be quickly detected
and a corrective action be taken proactively to avoid any downtime.
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 6
PACS make it possible to electronically store, manage, distribute, and view images. Funda-
mentally, these systems are a network of all image acquisition devices, display workstations,
and storage systems. They are made up of a broad range of technologies that enable digital
radiology and digital hospitals to perform tele-radiology, tele-medicine, and tele-surgery.
PACS have become more complex, encompassing systems that digitally acquire, convert,
interpret, transmit, and store medical images. Diagnostic images will be available anytime,
anywhere with little or no human intervention, making their distribution faster, easier, and
more reliable. Figure 5 illustrates the components of a typical PACS.
Environment
The core of the PACS is made up of high availability RAID storage and server clusters
running Windows, Unix, Linux, or a propriety operating system. These RAID storage and
server clusters are housed in racks in a computer / data room or data center environment.
Typically they draw less than 10 kVA, single-phase AC power at 120 V, 208V, or 230 VAC.
Very large systems may draw three-phase power.
Challenges
PACS need to be available on demand to the nurse, physician, clinician, or specialist
surgeon, providing latest imaging data of the patient under treatment. It needs to be highly
available, 7x24x365 and there is little tolerance to downtime. Since the RAID drives and
server clusters are confined to rack enclosures, handling their heat dissipation within the
racks often becomes a bigger challenge.
Best practices
• PACS should be protected with an N+1 redundant UPS system. This N+1 UPS system
not only protects the hardware but also protects the software from malfunctioning and
gracefully shuts down and reboots the operating system if needed, thereby preventing
a hard crash. The N+1 redundancy of the UPS system mirrors the redundancy of the
RAID storage drives and server clusters which are at the core of the PACS, providing
high availability. For smaller, simpler systems, a basic UPS can be provided.
• Often times, additional receptacles are needed for plugging in all required devices.
Rack based PDUs should be used to provide additional outlets. PDUs that can meas-
ure and display current, which can help prevent accidental overloading and shut down
of the PACS are recommended. PDUs that allow remote outlet control via the web are
desirable as they can help reboot a hung server or a storage drive efficiently.
Picture archiving
and
communication
systems (PACS)
Figure 5
Typical PACS
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 7
• At a minimum, surge suppression should be provided for LCD/CRT based passive
view-stations. For PC based active workstations running software applications, UPS
protection with graceful shut down and reboot capabilities is highly recommended.
• The PACS storage and servers should be housed in secured, lockable, rack enclo-
sures. These racks should be in a temperature controlled environment. The racks
housing PACS storage and servers are generally very dense physically and in terms of
power consumption. The rack doors should be perforated, allowing for maximum air-
flow. When power draw in the rack exceeds 4 kW, the rack should be placed in a hot
or cold aisle containment system. If this isn’t possible, ducted racks should be used to
capture hot return air and direct it back to the air conditioner. Alternatively, a row-
based cooling unit can be used to provide supplemental cooling capacity (see Figure
6). For more information on containment systems for existing data centers see White
Paper 153, Implementing Hot Aisle and Cold Aisle Containment in Existing Data Cen-
ters.
• A good management strategy involves the management of PACS servers, storage, and
their entire physical infrastructure including UPS, PDUs, batteries, and their critical en-
vironment (temperature and humidity). This will give early warning of any anomaly or
impending disaster so that corrective actions can be taken and prevent all avoidable
shutdowns.
RIS and HIS are server based systems running special software that make it possible to
store, monitor, manage, and distribute patient medical information. They help patients in
scheduling appointments, registration, and billing, and help hospitals in generating, maintain-
ing, and managing patient’s electronic medical records as well as generate workflow, work-
list, management reporting, and variety of other tasks. These RIS and HIS are really
becoming one large HIS and are integrated / networked with PACS as well as various other
modalities within the hospitals providing complete automation. By converting them into
“digital hospitals”, they can significantly improve patient-care, minimizing human errors,
saving lives, and reducing costs. Figure 7 shows a typical RIS / HIS and its subcomponents.
Environment
These systems are generally housed in a data center environment drawing 10 kW single-
phase 208 VAC or 230 VAC power on the lower side, to hundreds of kilowatts of three-phase
400 VAC or 480 VAC power on the higher side. The majority of data centers in hospitals
have a UPS with battery back-up, precision air conditioning units, and a back-up generator.
Challenges
RIS / HIS are the most important systems within the data center requiring longer runtime and
higher redundancy and availability then most other equipment. Since these systems are
Figure 6
Example of ducted rack air
containment (left)
(Schneider Electric Vertical Exhaust
Duct shown)
Example of row-based cooling
unit (right)
(Schneider Electric InRow RC shown)
Radiology
information
systems (RIS)
and hospital
information
systems (HIS)
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 8
merging to form one big HIS on which the entire hospital depends for normal functioning,
their availability requirements are generally 99.999% (five nines) or higher which translates to
average unplanned downtime of 5 minutes per year or less. Additionally these systems may
be located in high rise buildings and attention should be paid to the floor load (weight)
handling capacity, elevator hauling capacity, door heights and widths to ensure that the
physical infrastructure elements like UPS, batteries, and air-conditioning can be rolled in to
their planned positions.
Hospital Wards
Transcriptionists
Reception
Management
Reporting
Medical
Technical
Department
Clinical
Departments
Best practices
• The physical infrastructure supporting the RIS / HIS should provide highest levels of
redundancy while minimizing the total cost of ownership. An N+1 redundant UPS with
automatic and manual bypass is very common and often times it is extended to the
generator as well as the precision air-conditioning systems to ensure the highest levels
of availability. The entire infrastructure should be scalable to allow for future expan-
sion, be manageable like the other IT equipment, and be serviceable to reduce mean
time to recover. An example of such a system is the Schneider Electric InfraStruxure
and is shown in Figures 8. All of these characteristics contribute to the overall availa-
bility of the system.
• Servers and systems requiring the highest levels of availability should be identified and
grouped so that they can be provided with longer runtime and higher levels of redun-
dancy in a separate area, and in separate racks within the data center. This concept of
“targeted availability” helps increase availability of business critical systems without
having to incur a large capital expense for the entire data center. Higher levels of re-
dundancy like dual feeds with dual generators and dual N+1 UPS with dual power
paths all the way to the rack should be considered for highly-critical data centers and
networks.
• PDUs should be able to measure and display current, which can help prevent acci-
dental overloading and shutdown of the RIS / HIS. PDUs that allow remote outlet con-
Figure 7
Typical RIS / HIS system
Figure 8
Schneider Electric
InfraStruxure™
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 9
trol via the web are desirable for rapidly rebooting a hung server or a storage drive.
Isolation transformers should be used wherever required and mandated by local laws.
• Precision air conditioning equipment should have the capability to allow for expansion.
Redundant air conditioning units should be considered for higher availability. For high
power density racks (>4 kW / rack) additional air distribution and air containment
should be used to avoid hot spots. For more information on cooling best practices refer
to White Paper 49, Avoidable Mistakes that Compromise Cooling Performance in Data
Centers and Network Rooms.
Medical imaging and diagnostic equipment are typically connected to a network. Modalities
like CTs and MRIs get connected to PACS which are connected to RIS and HIS which in turn
are connected to the hospital intranets and extranets. The wiring closets or IDFs, as shown
in Figure 9, play a very vital role in ensuring the connectivity of this equipment and the
availability of the network, 7x24x365. Wiring closets comprise of layer 2 and layer 3 access
and distribution switches, hubs, routers, patch panels, UPS systems with a battery back-up
as well as any other miscellaneous telecommunications equipment mounted generally in a
two post rack. IDFs or wiring closets may also supply power over Ethernet (PoE) to net-
worked devices likes IP phones, web/security camera and any other devices drawing power
up to 15 W. This imposes a lot more challenges on the power and cooling requirements in
the closets.
Environment
These IDFs or wiring closets are typically hidden in some remote location of the building with
little or no ventilation and illumination. Legacy telecommunication networks typically used
wiring closets mainly for punch-down blocks, patch panels, and a few small stackable hubs or
switches. However, networking equipment that supplies power over Ethernet uses and
dissipates considerably more power. These new switches support data, voice, and video,
generally 19” rack mount type, and have varying air flow patterns depending on the manufac-
turers (i.e. side to side vs. front to back). A typical IDF will house 1-3 racks worth of equip-
ment and draw 500 W to 4000 W of single phase AC power or more. Two post racks have
largely been replaced with four post racks since newer equipment is getting heavier and
deeper.
Wiring closets or
intermediate
distribution
frame (IDF)
Patch Panel
Midspan Power
Supply
Network Telephony
System
Network Switches
Uninterrutible
Power
Supply
Figure 9
IDF (wiring closet)
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 10
Challenges
While deploying PACS, RIS, HIS, or new modalities that are being networked, these wiring
closets or IDFs need the most attention in terms of power and cooling. It can be a challenge
to ensure the right type of receptacles (i.e. L5-20, L5-30, L6-20, IEC 320 C19, and IEC 320
C13) and the right amount of power with the right circuit breaker protection to all the network-
ing equipment. Cooling and airflow are often a bigger but ignored problem to address in
these wiring closets.
Best practices
All equipment in the IDF should be protected by a UPS system. The selection of UPS should
be based on:
• The total power required in Watts
• The run time required in minutes
• The level of redundancy or fault tolerance desired
• The voltages and receptacles required
The UPS system is sized by taking the sum of the Watt ratings of the loads. A common rack-
mount UPS (i.e. Smart-UPS™) will provide approximately four nines (99.99%) of power
availability, while an N+1 redundant, UPS with built in bypass (i.e. Symmetra RM) with one
hour runtime will provide approximately five nines (99.999%), which may be sufficient for
most applications. See Appendix of White Paper 69, Power and Cooling for VoIP and IP
Telephony Applications for details on this availability analysis. UPS products are available
with battery packs to provide different durations of run time.
Identify the plugs and receptacles required for all the equipment including the UPS in the
wiring closet. Ideally all of the equipment should be plugged directly into the back of the UPS
or the transformer, and the use of additional outlet strips or rack PDUs should be avoided.
However, if there are many devices, it may not be practical and a rack PDU should be used.
In that case a high-grade rack PDU specifically designed for the purpose should be used.
The PDU should have enough receptacles to plug all the current equipment with some spares
for future needs. PDUs with a meter displaying the current power consumption are preferred
as they reduce human error like accidental overloading and resultant load drops. For the
correct selection of the appropriate UPS model meeting the required power level, redundan-
cy, voltage, and run time, the process is simplified by using a UPS selector such as the APC
by Schneider Electric UPS Selector. This system has power data for all popular switches,
servers, and storage devices, which avoids the need to collect this data. In systems like this,
the choice of configuring a UPS will provide various receptacle options.
To ensure continuous operations of the equipment in the wiring closet, 7x24x365, cooling and
airflow issues must be identified and addressed. The problem of heat dissipation and need
for supplemental air-conditioning is most pronounced in the wiring closets which have no
vents. Power dissipation in the wiring closet should be calculated to decide on a cost
effective way to solve the problem (see Table 1 & Table 2 in White Paper 69, Power and
Cooling for VoIP and IP Telephony Applications for details).
Finally, environmental monitoring (i.e. temperature and humidity) within these wiring closets
is highly recommended as it will help flag any abnormal conditions, allow for enough time to
take proactive measures, and avoid downtime.
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 11
To ensure high availability and reliability to medical imaging and diagnostic equipment,
including PACS, RIS, HIS, modalities, and their network, special attention must be paid to
their physical infrastructure. The biggest challenges are in terms of power, cooling, physical
space, management, and services. Providing UPS protection to all such devices protects the
hardware, prevents the software from unwarranted crashes, and increases their availability
significantly. Cooling is a special problem for bigger floor mounted modalities, high density
storage and servers for PACS as well as RIS / HIS and hospital wiring closets. In some
cases, a building’s HVAC system along with proper ducting, ventilation, and airflow may be
sufficient. However in many situations, additional cooling in the form of precision air condi-
tioning is required. Companies like Schneider Electric have dedicated team of systems
engineers, power protection specialists, and availability consultants who can help in doing
assessments and audits of data center physical infrastructure and provide detailed actionable
reports on improving overall system reliability and availability while minimizing the total cost
of ownership.
Conclusion
Special thanks to Viswas Purani for authoring the original content of this white paper.
Acknowledgements
Power Protection for Digital Medical Imaging and Diagnostic Equipment
Schneider Electric – Data Center Science Center Rev 0 12
The Different Types of UPS Systems
White Paper 1
Avoiding costs from Oversizing Data Center and
Network Room Infrastructure
White Paper 37
Dynamic Power Variations in Data Centers and Network Rooms
White Paper 43
Avoidable Mistakes that Compromise Cooling
Performance in Data Centers and Network Rooms
White Paper 49
Power and Cooling for VoIP and IP Telephony Applications
White Paper 69
Implementing Hot Aisle and Cold Aisle Containment
in Existing Data Centers
White Paper 153
IEEE White book “IEEE Recommended Practice for Electric Systems in Health Care Facili-
ties” IEEE std. 602-1996
References
Resources
Browse all
white papers
whitepapers.apc.com
tools.apc.com
Browse all
TradeOff Tools™
©2014SchneiderElectric.Allrightsreserved.
For feedback and comments about the content of this white paper:
Data Center Science Center
dcsc@schneider-electric.com
If you are a customer and have questions specific to your data center project:
Contact your Schneider Electric representative at
www.apc.com/support/contact/index.cfm
Contact us

Más contenido relacionado

La actualidad más candente

Types of Prefabricated Modular Data Centers
Types of Prefabricated Modular Data CentersTypes of Prefabricated Modular Data Centers
Types of Prefabricated Modular Data CentersSchneider Electric
 
Industrial sawmill solution Wood Mizer
Industrial sawmill solution Wood MizerIndustrial sawmill solution Wood Mizer
Industrial sawmill solution Wood MizerSchneider Electric
 
Integrated Control and Safety - Assessing the Benefits; Weighing the Risks
Integrated Control and Safety - Assessing the Benefits; Weighing the RisksIntegrated Control and Safety - Assessing the Benefits; Weighing the Risks
Integrated Control and Safety - Assessing the Benefits; Weighing the RisksSchneider Electric
 
Dtech 2015 the distribution management system network model
Dtech 2015  the distribution management system network modelDtech 2015  the distribution management system network model
Dtech 2015 the distribution management system network modelSchneider Electric
 
Asset Management - what are some of your top priorties?
Asset Management - what are some of your top priorties?Asset Management - what are some of your top priorties?
Asset Management - what are some of your top priorties?Schneider Electric
 
How the Convergence of IT and OT Enables Smart Grid Development
How the Convergence of IT and OT Enables Smart Grid DevelopmentHow the Convergence of IT and OT Enables Smart Grid Development
How the Convergence of IT and OT Enables Smart Grid DevelopmentSchneider Electric
 
Advantages of a modern DCS in coal handling
Advantages of a modern DCS in coal handlingAdvantages of a modern DCS in coal handling
Advantages of a modern DCS in coal handlingSchneider Electric
 
Energy efficiency data center overview
Energy efficiency data center overviewEnergy efficiency data center overview
Energy efficiency data center overviewSchneider Electric
 
Modicon M580 MUCH MORE then a PLC…the first ePAC!
Modicon M580 MUCH MORE then a PLC…the first ePAC!Modicon M580 MUCH MORE then a PLC…the first ePAC!
Modicon M580 MUCH MORE then a PLC…the first ePAC!Schneider Electric
 
A framework for converting hotel guestroom energy management into ROI
A framework for converting hotel guestroom energy management into ROIA framework for converting hotel guestroom energy management into ROI
A framework for converting hotel guestroom energy management into ROISchneider Electric
 
New Approach for Intelligent Motor Control Centers
New Approach for Intelligent Motor Control CentersNew Approach for Intelligent Motor Control Centers
New Approach for Intelligent Motor Control CentersSchneider Electric
 
Field Data Gathering Services — A Cloud-Based Approach
Field Data Gathering Services — A Cloud-Based ApproachField Data Gathering Services — A Cloud-Based Approach
Field Data Gathering Services — A Cloud-Based ApproachSchneider Electric
 
Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...
Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...
Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...Schneider Electric
 
Advanced Weather Systems for Mining Operations
Advanced Weather Systems for Mining OperationsAdvanced Weather Systems for Mining Operations
Advanced Weather Systems for Mining OperationsSchneider Electric
 
Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...
Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...
Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...Schneider Electric
 
Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...
Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...
Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...Schneider Electric
 
[Webinar Presentation] Best Practices for IT/OT Convergence
[Webinar Presentation] Best Practices for IT/OT Convergence[Webinar Presentation] Best Practices for IT/OT Convergence
[Webinar Presentation] Best Practices for IT/OT ConvergenceSchneider Electric
 

La actualidad más candente (20)

Types of Prefabricated Modular Data Centers
Types of Prefabricated Modular Data CentersTypes of Prefabricated Modular Data Centers
Types of Prefabricated Modular Data Centers
 
Industrial sawmill solution Wood Mizer
Industrial sawmill solution Wood MizerIndustrial sawmill solution Wood Mizer
Industrial sawmill solution Wood Mizer
 
Integrated Control and Safety - Assessing the Benefits; Weighing the Risks
Integrated Control and Safety - Assessing the Benefits; Weighing the RisksIntegrated Control and Safety - Assessing the Benefits; Weighing the Risks
Integrated Control and Safety - Assessing the Benefits; Weighing the Risks
 
Dtech 2015 the distribution management system network model
Dtech 2015  the distribution management system network modelDtech 2015  the distribution management system network model
Dtech 2015 the distribution management system network model
 
Asset Management - what are some of your top priorties?
Asset Management - what are some of your top priorties?Asset Management - what are some of your top priorties?
Asset Management - what are some of your top priorties?
 
How the Convergence of IT and OT Enables Smart Grid Development
How the Convergence of IT and OT Enables Smart Grid DevelopmentHow the Convergence of IT and OT Enables Smart Grid Development
How the Convergence of IT and OT Enables Smart Grid Development
 
Advantages of a modern DCS in coal handling
Advantages of a modern DCS in coal handlingAdvantages of a modern DCS in coal handling
Advantages of a modern DCS in coal handling
 
Energy efficiency data center overview
Energy efficiency data center overviewEnergy efficiency data center overview
Energy efficiency data center overview
 
Modicon M580 MUCH MORE then a PLC…the first ePAC!
Modicon M580 MUCH MORE then a PLC…the first ePAC!Modicon M580 MUCH MORE then a PLC…the first ePAC!
Modicon M580 MUCH MORE then a PLC…the first ePAC!
 
A framework for converting hotel guestroom energy management into ROI
A framework for converting hotel guestroom energy management into ROIA framework for converting hotel guestroom energy management into ROI
A framework for converting hotel guestroom energy management into ROI
 
Data centers on the Edge
Data centers on the EdgeData centers on the Edge
Data centers on the Edge
 
New Approach for Intelligent Motor Control Centers
New Approach for Intelligent Motor Control CentersNew Approach for Intelligent Motor Control Centers
New Approach for Intelligent Motor Control Centers
 
Field Data Gathering Services — A Cloud-Based Approach
Field Data Gathering Services — A Cloud-Based ApproachField Data Gathering Services — A Cloud-Based Approach
Field Data Gathering Services — A Cloud-Based Approach
 
Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...
Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...
Impact of IEC 61508 Standards on Intelligent Electrial Networks and Safety Im...
 
Advanced Weather Systems for Mining Operations
Advanced Weather Systems for Mining OperationsAdvanced Weather Systems for Mining Operations
Advanced Weather Systems for Mining Operations
 
Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...
Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...
Virtualization and Cloud Computing: Optimized Power, Cooling, and Management ...
 
AccuSine pfv+
AccuSine pfv+ AccuSine pfv+
AccuSine pfv+
 
Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...
Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...
Practical Options for Deploying IT Equipment in Small Server Rooms and Branch...
 
[Webinar Presentation] Best Practices for IT/OT Convergence
[Webinar Presentation] Best Practices for IT/OT Convergence[Webinar Presentation] Best Practices for IT/OT Convergence
[Webinar Presentation] Best Practices for IT/OT Convergence
 
Holistic Efficiency
Holistic EfficiencyHolistic Efficiency
Holistic Efficiency
 

Destacado

How Schneider Electric sees Ethernet in the Industrial Environment - Part 1
How Schneider Electric sees Ethernet in the Industrial Environment - Part 1How Schneider Electric sees Ethernet in the Industrial Environment - Part 1
How Schneider Electric sees Ethernet in the Industrial Environment - Part 1Schneider Electric
 
Hannover High End Utility Metering
Hannover High End Utility MeteringHannover High End Utility Metering
Hannover High End Utility MeteringSchneider Electric
 
Schneider Electric
Schneider ElectricSchneider Electric
Schneider Electricsudhakar_hr
 
Short-Circuit Current Ratings and AC Drives
Short-Circuit Current Ratings and AC DrivesShort-Circuit Current Ratings and AC Drives
Short-Circuit Current Ratings and AC DrivesSchneider Electric
 
Schneider Electric Facilities Management Survey
Schneider Electric Facilities Management SurveySchneider Electric Facilities Management Survey
Schneider Electric Facilities Management SurveySchneider Electric
 
Industrial Energy Management and the Emerging ISO 50001 Standard
Industrial Energy Management and the Emerging ISO 50001 StandardIndustrial Energy Management and the Emerging ISO 50001 Standard
Industrial Energy Management and the Emerging ISO 50001 StandardSchneider Electric
 
Energy management & audit
Energy management & auditEnergy management & audit
Energy management & auditPINAKI ROY
 

Destacado (9)

How Schneider Electric sees Ethernet in the Industrial Environment - Part 1
How Schneider Electric sees Ethernet in the Industrial Environment - Part 1How Schneider Electric sees Ethernet in the Industrial Environment - Part 1
How Schneider Electric sees Ethernet in the Industrial Environment - Part 1
 
Hannover High End Utility Metering
Hannover High End Utility MeteringHannover High End Utility Metering
Hannover High End Utility Metering
 
Schneider Electric
Schneider ElectricSchneider Electric
Schneider Electric
 
Short-Circuit Current Ratings and AC Drives
Short-Circuit Current Ratings and AC DrivesShort-Circuit Current Ratings and AC Drives
Short-Circuit Current Ratings and AC Drives
 
Schneider Electric Facilities Management Survey
Schneider Electric Facilities Management SurveySchneider Electric Facilities Management Survey
Schneider Electric Facilities Management Survey
 
Paul Stride - Electrical earth safety
Paul Stride  - Electrical earth safetyPaul Stride  - Electrical earth safety
Paul Stride - Electrical earth safety
 
Smart City - Cornerstone of Efficiency
Smart City - Cornerstone of EfficiencySmart City - Cornerstone of Efficiency
Smart City - Cornerstone of Efficiency
 
Industrial Energy Management and the Emerging ISO 50001 Standard
Industrial Energy Management and the Emerging ISO 50001 StandardIndustrial Energy Management and the Emerging ISO 50001 Standard
Industrial Energy Management and the Emerging ISO 50001 Standard
 
Energy management & audit
Energy management & auditEnergy management & audit
Energy management & audit
 

Similar a Power Protection for Digital Medical Imaging and Diagnostic Equipment

Security in Body Sensor Networks for Healthcare applications
Security in Body Sensor Networks for Healthcare applicationsSecurity in Body Sensor Networks for Healthcare applications
Security in Body Sensor Networks for Healthcare applicationsIOSR Journals
 
Wireless Sensor Network for Patient Health Monitoring System
Wireless Sensor Network for Patient Health Monitoring SystemWireless Sensor Network for Patient Health Monitoring System
Wireless Sensor Network for Patient Health Monitoring SystemIRJET Journal
 
Wireless Sensor Network: an emerging entrant in Healthcare
Wireless Sensor Network: an emerging entrant in HealthcareWireless Sensor Network: an emerging entrant in Healthcare
Wireless Sensor Network: an emerging entrant in HealthcareIOSR Journals
 
Telemedicine technology positioning
Telemedicine technology positioningTelemedicine technology positioning
Telemedicine technology positioningtibisay_hernandez
 
IRJET- Hiding Sensitive Medical Data using Encryption
IRJET- Hiding Sensitive Medical Data using EncryptionIRJET- Hiding Sensitive Medical Data using Encryption
IRJET- Hiding Sensitive Medical Data using EncryptionIRJET Journal
 
Iaetsd io t based advanced smart health care system
Iaetsd io t based advanced smart health care systemIaetsd io t based advanced smart health care system
Iaetsd io t based advanced smart health care systemIaetsd Iaetsd
 
A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...
A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...
A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...ijait
 
Power Quality for Healthcare Facilities
Power Quality for Healthcare FacilitiesPower Quality for Healthcare Facilities
Power Quality for Healthcare FacilitiesLeonardo ENERGY
 
security and privacy for medical implantable devices
security and privacy for medical implantable devicessecurity and privacy for medical implantable devices
security and privacy for medical implantable devicesAjay Ohri
 
A New Real Time Clinical Decision Support System Using Machine Learning for C...
A New Real Time Clinical Decision Support System Using Machine Learning for C...A New Real Time Clinical Decision Support System Using Machine Learning for C...
A New Real Time Clinical Decision Support System Using Machine Learning for C...IRJET Journal
 
Energy Aware Computing in Sensor Networks
Energy Aware Computing in Sensor NetworksEnergy Aware Computing in Sensor Networks
Energy Aware Computing in Sensor Networkspaperpublications3
 
Wide area protection-research_in_the_smart_grid
Wide area protection-research_in_the_smart_gridWide area protection-research_in_the_smart_grid
Wide area protection-research_in_the_smart_gridAlaa Eladl
 
Dynamic Medical Machine
Dynamic Medical MachineDynamic Medical Machine
Dynamic Medical MachineIRJET Journal
 
Computers in Medical field
Computers in Medical fieldComputers in Medical field
Computers in Medical fieldAvinash
 

Similar a Power Protection for Digital Medical Imaging and Diagnostic Equipment (20)

Components And Workflow Of A Digital Radiology Department
Components And Workflow Of A Digital Radiology DepartmentComponents And Workflow Of A Digital Radiology Department
Components And Workflow Of A Digital Radiology Department
 
Secret key generation
Secret key generationSecret key generation
Secret key generation
 
Security in Body Sensor Networks for Healthcare applications
Security in Body Sensor Networks for Healthcare applicationsSecurity in Body Sensor Networks for Healthcare applications
Security in Body Sensor Networks for Healthcare applications
 
Wireless Sensor Network for Patient Health Monitoring System
Wireless Sensor Network for Patient Health Monitoring SystemWireless Sensor Network for Patient Health Monitoring System
Wireless Sensor Network for Patient Health Monitoring System
 
Wireless Sensor Network: an emerging entrant in Healthcare
Wireless Sensor Network: an emerging entrant in HealthcareWireless Sensor Network: an emerging entrant in Healthcare
Wireless Sensor Network: an emerging entrant in Healthcare
 
Telemedicine technology positioning
Telemedicine technology positioningTelemedicine technology positioning
Telemedicine technology positioning
 
IRJET- Hiding Sensitive Medical Data using Encryption
IRJET- Hiding Sensitive Medical Data using EncryptionIRJET- Hiding Sensitive Medical Data using Encryption
IRJET- Hiding Sensitive Medical Data using Encryption
 
Iaetsd io t based advanced smart health care system
Iaetsd io t based advanced smart health care systemIaetsd io t based advanced smart health care system
Iaetsd io t based advanced smart health care system
 
A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...
A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...
A FRAMEWORK FOR THE INTERCONNECTION OF CONTROLLER AREA NETWORK (CAN) BASED CR...
 
Bender UK Turnkey Healthcare Solutions
Bender UK Turnkey Healthcare SolutionsBender UK Turnkey Healthcare Solutions
Bender UK Turnkey Healthcare Solutions
 
C1031724
C1031724C1031724
C1031724
 
Power Quality for Healthcare Facilities
Power Quality for Healthcare FacilitiesPower Quality for Healthcare Facilities
Power Quality for Healthcare Facilities
 
Wireless Sensor Networks for Monitoring Physiological Signals of Multiple Pat...
Wireless Sensor Networks for Monitoring Physiological Signals of Multiple Pat...Wireless Sensor Networks for Monitoring Physiological Signals of Multiple Pat...
Wireless Sensor Networks for Monitoring Physiological Signals of Multiple Pat...
 
security and privacy for medical implantable devices
security and privacy for medical implantable devicessecurity and privacy for medical implantable devices
security and privacy for medical implantable devices
 
A New Real Time Clinical Decision Support System Using Machine Learning for C...
A New Real Time Clinical Decision Support System Using Machine Learning for C...A New Real Time Clinical Decision Support System Using Machine Learning for C...
A New Real Time Clinical Decision Support System Using Machine Learning for C...
 
Energy Aware Computing in Sensor Networks
Energy Aware Computing in Sensor NetworksEnergy Aware Computing in Sensor Networks
Energy Aware Computing in Sensor Networks
 
Wide area protection-research_in_the_smart_grid
Wide area protection-research_in_the_smart_gridWide area protection-research_in_the_smart_grid
Wide area protection-research_in_the_smart_grid
 
Dynamic Medical Machine
Dynamic Medical MachineDynamic Medical Machine
Dynamic Medical Machine
 
Computers in Medical field
Computers in Medical fieldComputers in Medical field
Computers in Medical field
 
paper1.pdf
paper1.pdfpaper1.pdf
paper1.pdf
 

Más de Schneider Electric

Secure Power Design Considerations
Secure Power Design ConsiderationsSecure Power Design Considerations
Secure Power Design ConsiderationsSchneider Electric
 
Digital International Colo Club: Attracting Investors
Digital International Colo Club: Attracting InvestorsDigital International Colo Club: Attracting Investors
Digital International Colo Club: Attracting InvestorsSchneider Electric
 
32 phaseo power supplies and transformers briefing
32 phaseo power supplies and transformers briefing 32 phaseo power supplies and transformers briefing
32 phaseo power supplies and transformers briefing Schneider Electric
 
Key Industry Trends, M&A Valuation Trends
Key Industry Trends, M&A Valuation TrendsKey Industry Trends, M&A Valuation Trends
Key Industry Trends, M&A Valuation TrendsSchneider Electric
 
EcoStruxure™ for Cloud & Service Providers
 EcoStruxure™ for Cloud & Service Providers EcoStruxure™ for Cloud & Service Providers
EcoStruxure™ for Cloud & Service ProvidersSchneider Electric
 
Zelio Time Electronic Relay Briefing
Zelio Time Electronic Relay BriefingZelio Time Electronic Relay Briefing
Zelio Time Electronic Relay BriefingSchneider Electric
 
Spacial, Thalassa, ClimaSys Universal enclosures Briefing
Spacial, Thalassa, ClimaSys Universal enclosures BriefingSpacial, Thalassa, ClimaSys Universal enclosures Briefing
Spacial, Thalassa, ClimaSys Universal enclosures BriefingSchneider Electric
 
Relay Control Zelio SSR Briefing
Relay Control Zelio SSR BriefingRelay Control Zelio SSR Briefing
Relay Control Zelio SSR BriefingSchneider Electric
 
Magelis HMI, iPC and software Briefing
Magelis HMI, iPC and software BriefingMagelis HMI, iPC and software Briefing
Magelis HMI, iPC and software BriefingSchneider Electric
 
Where will the next 80% improvement in data center performance come from?
Where will the next 80% improvement in data center performance come from?Where will the next 80% improvement in data center performance come from?
Where will the next 80% improvement in data center performance come from?Schneider Electric
 
EcoStruxure for Intuitive Industries
EcoStruxure for Intuitive IndustriesEcoStruxure for Intuitive Industries
EcoStruxure for Intuitive IndustriesSchneider Electric
 
Systems Integrator Alliance Program 2017
Systems Integrator Alliance Program 2017Systems Integrator Alliance Program 2017
Systems Integrator Alliance Program 2017Schneider Electric
 
EcoStruxure, IIoT-enabled architecture, delivering value in key segments.
EcoStruxure, IIoT-enabled architecture, delivering value in key segments.EcoStruxure, IIoT-enabled architecture, delivering value in key segments.
EcoStruxure, IIoT-enabled architecture, delivering value in key segments.Schneider Electric
 
It's time to modernize your industrial controls with Modicon M580
It's time to modernize your industrial controls with Modicon M580It's time to modernize your industrial controls with Modicon M580
It's time to modernize your industrial controls with Modicon M580Schneider Electric
 
A Practical Guide to Ensuring Business Continuity and High Performance in Hea...
A Practical Guide to Ensuring Business Continuity and High Performance in Hea...A Practical Guide to Ensuring Business Continuity and High Performance in Hea...
A Practical Guide to Ensuring Business Continuity and High Performance in Hea...Schneider Electric
 
Connected Services Study – Facility Managers Respond to IoT
Connected Services Study – Facility Managers Respond to IoTConnected Services Study – Facility Managers Respond to IoT
Connected Services Study – Facility Managers Respond to IoTSchneider Electric
 
Telemecanqiue Cabling and Accessories Briefing
Telemecanqiue Cabling and Accessories BriefingTelemecanqiue Cabling and Accessories Briefing
Telemecanqiue Cabling and Accessories BriefingSchneider Electric
 
Telemecanique Photoelectric Sensors Briefing
Telemecanique Photoelectric Sensors BriefingTelemecanique Photoelectric Sensors Briefing
Telemecanique Photoelectric Sensors BriefingSchneider Electric
 
Telemecanique Limit Switches Briefing
Telemecanique Limit Switches BriefingTelemecanique Limit Switches Briefing
Telemecanique Limit Switches BriefingSchneider Electric
 

Más de Schneider Electric (20)

Secure Power Design Considerations
Secure Power Design ConsiderationsSecure Power Design Considerations
Secure Power Design Considerations
 
Digital International Colo Club: Attracting Investors
Digital International Colo Club: Attracting InvestorsDigital International Colo Club: Attracting Investors
Digital International Colo Club: Attracting Investors
 
32 phaseo power supplies and transformers briefing
32 phaseo power supplies and transformers briefing 32 phaseo power supplies and transformers briefing
32 phaseo power supplies and transformers briefing
 
Key Industry Trends, M&A Valuation Trends
Key Industry Trends, M&A Valuation TrendsKey Industry Trends, M&A Valuation Trends
Key Industry Trends, M&A Valuation Trends
 
EcoStruxure™ for Cloud & Service Providers
 EcoStruxure™ for Cloud & Service Providers EcoStruxure™ for Cloud & Service Providers
EcoStruxure™ for Cloud & Service Providers
 
Magelis Basic HMI Briefing
Magelis Basic HMI Briefing Magelis Basic HMI Briefing
Magelis Basic HMI Briefing
 
Zelio Time Electronic Relay Briefing
Zelio Time Electronic Relay BriefingZelio Time Electronic Relay Briefing
Zelio Time Electronic Relay Briefing
 
Spacial, Thalassa, ClimaSys Universal enclosures Briefing
Spacial, Thalassa, ClimaSys Universal enclosures BriefingSpacial, Thalassa, ClimaSys Universal enclosures Briefing
Spacial, Thalassa, ClimaSys Universal enclosures Briefing
 
Relay Control Zelio SSR Briefing
Relay Control Zelio SSR BriefingRelay Control Zelio SSR Briefing
Relay Control Zelio SSR Briefing
 
Magelis HMI, iPC and software Briefing
Magelis HMI, iPC and software BriefingMagelis HMI, iPC and software Briefing
Magelis HMI, iPC and software Briefing
 
Where will the next 80% improvement in data center performance come from?
Where will the next 80% improvement in data center performance come from?Where will the next 80% improvement in data center performance come from?
Where will the next 80% improvement in data center performance come from?
 
EcoStruxure for Intuitive Industries
EcoStruxure for Intuitive IndustriesEcoStruxure for Intuitive Industries
EcoStruxure for Intuitive Industries
 
Systems Integrator Alliance Program 2017
Systems Integrator Alliance Program 2017Systems Integrator Alliance Program 2017
Systems Integrator Alliance Program 2017
 
EcoStruxure, IIoT-enabled architecture, delivering value in key segments.
EcoStruxure, IIoT-enabled architecture, delivering value in key segments.EcoStruxure, IIoT-enabled architecture, delivering value in key segments.
EcoStruxure, IIoT-enabled architecture, delivering value in key segments.
 
It's time to modernize your industrial controls with Modicon M580
It's time to modernize your industrial controls with Modicon M580It's time to modernize your industrial controls with Modicon M580
It's time to modernize your industrial controls with Modicon M580
 
A Practical Guide to Ensuring Business Continuity and High Performance in Hea...
A Practical Guide to Ensuring Business Continuity and High Performance in Hea...A Practical Guide to Ensuring Business Continuity and High Performance in Hea...
A Practical Guide to Ensuring Business Continuity and High Performance in Hea...
 
Connected Services Study – Facility Managers Respond to IoT
Connected Services Study – Facility Managers Respond to IoTConnected Services Study – Facility Managers Respond to IoT
Connected Services Study – Facility Managers Respond to IoT
 
Telemecanqiue Cabling and Accessories Briefing
Telemecanqiue Cabling and Accessories BriefingTelemecanqiue Cabling and Accessories Briefing
Telemecanqiue Cabling and Accessories Briefing
 
Telemecanique Photoelectric Sensors Briefing
Telemecanique Photoelectric Sensors BriefingTelemecanique Photoelectric Sensors Briefing
Telemecanique Photoelectric Sensors Briefing
 
Telemecanique Limit Switches Briefing
Telemecanique Limit Switches BriefingTelemecanique Limit Switches Briefing
Telemecanique Limit Switches Briefing
 

Último

hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...delhimodelshub1
 
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy GirlsRussian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girlsddev2574
 
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...delhimodelshub1
 
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service HyderabadVIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...scanFOAM
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Servicenarwatsonia7
 
Call Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any TimeCall Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any Timedelhimodelshub1
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed RuleShelby Lewis
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxAyushGupta813444
 
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service HyderabadCall Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service GoaRussian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goanarwatsonia7
 
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...narwatsonia7
 
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...narwatsonia7
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowHyderabad Call Girls Services
 
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near MeBook Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 

Último (20)

hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
 
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy GirlsRussian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
 
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
 
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
 
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service HyderabadVIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
 
Call Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any TimeCall Girls Uppal 7001305949 all area service COD available Any Time
Call Girls Uppal 7001305949 all area service COD available Any Time
 
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
Model Call Girl in Subhash Nagar Delhi reach out to us at 🔝9953056974🔝
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptx
 
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service HyderabadCall Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
 
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service GoaRussian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
 
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
 
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
 
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near MeBook Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
Book Call Girls in Hosur - 7001305949 | 24x7 Service Available Near Me
 
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 

Power Protection for Digital Medical Imaging and Diagnostic Equipment

  • 1. Power Protection for Digital Medical Imaging and Diagnostic Equipment Revision 2 by Viswas Purani White Paper 86 by Schneider Electric White Papers are now part of the Schneider Electric white paper library produced by Schneider Electric’s Data Center Science Center DCSC@Schneider-Electric.com Medical imaging and diagnostic equipment (MIDE) is increasingly being networked to Picture Archiving and Communications Systems (PACS), Radiology Infor- mation Systems (RIS), Hospital Information Systems (HIS), and getting connected to the hospital intranet as well as the Internet. Failing to implement the neces- sary physical infrastructure can result in unexpected downtime, and safety and compliance issues, which translates into lost revenue and exposure to expensive litigations, negatively affecting the bottom line. This paper explains how to plan for physical infrastructure when deploying medical imaging and diagnostic equipment, with emphasis on power and cooling. Executive summary
  • 2. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 2 The proliferation of information technology (IT) and other high technologies into medical imaging and diagnostic equipment over the years has resulted in the evolution of powerful new devices in the field of diagnostics and interventional radiology. The information generat- ed and carried by these images is crucial to the treatment in cardiology, neurology, oncology, gynecology & obstetrics, orthopedics, surgery, and pulmonary medicine. These new developments have helped in the early detection and treatment of diseases and significantly improved patient care. A typical medical imaging and diagnostic equipment network is illustrated in Figure 1. This equipment can be broken down into five broad sub-categories: 1. Modalities that capture or generate the images 2. Picture Archiving and Communications Systems (PACS) that store the generated images and make them available to the physicians and nurses for diagnosis and treatment 3. Radiology Information Systems (RIS) and Hospital Information Systems (HIS) that monitor and manage the work flow of radiology departments and entire hospitals, all the way from the patient check-in, to scheduling and billing, to generating electronic medical records and management reporting 4. Computed Radiography (CR) that helps convert films to digital images on cassettes or Digital Radiography (DR) that provides cassette-less digital images 5. Laser printers that print films when required and other peripherals CT, MRI, etc. RIS Viewing Workstations Modality HIS Consulting Cardiologist Consulting Physician Internet VPN Storage - PACS Modalities are high-tech medical imaging systems including Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Positron Emission Tomography (PET), Ultrasound (US), and Electro-Cardiogram (ECG). They get connected to a PACS and to RIS/HIS through Local Area Networks (LANs), and/or Wireless Local Area Networks (WLANs), or Wide Area Networks (WANs). PACS may have their own Storage Area Networks (SANs) or Network Attached Storage (NAS) and the RIS and HIS may be made of several clusters of servers and a number of workstations distributed in different departments of the hospitals. Because of their numerous benefits combined with the enormous pressure on the hospitals to improve quality of care, reduce errors, comply with federal regulations like HIPAA, and simultaneously cut cost, adoption of all these technologies is inevitable, converting the traditional hospital into a “digital enterprise”. Introduction Figure 1 Typical MIDE network
  • 3. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 3 The backbone of this new digital hospital is a network made up of different modalities, PACS and RIS/HIS, CR/DR, printers, and peripherals. This highly complex network and its compo- nents has to comply with relevant standards like Digital Imaging and Communications in Medicine (DICOM), Health Language Seven (HL7), Underwriters Laboratory (UL), Federal Communications Commission (FCC), National Electrical Code (NEC), and other applicable local and national codes such as BS7671:2001 (U.K.), NFC15-100 (France), and VDE (Germany), as well as international such as CEI IEC 60364. This imposes a huge challenge to the IT and the facilities manager to provide the right physical infrastructure. This physical infrastructure has to be reliable, scalable, highly available, and manageable. It consists of: 1. Power systems such as uninterruptible power supplies (UPSs), power distribution units (PDUs), isolation transformers, and generators to provide uninterrupted, condi- tioned, clean power to the critical loads 2. Precision cooling systems that provide optimal environment by regulating temperature and humidity 3. Racks that house the critical network equipment like servers, switches, routers, and gateways 4. Physical security and fire protection systems 5. Cabling to interconnect equipment 6. Management systems to monitor and manage these systems, locally as well as re- motely to ensure their satisfactory operation 7x24x365 7. Services to design, deliver, install, commission, operate and maintain these systems Special attention should be given to the hospital wiring closets which allow networking of the modalities to PACS and RIS/HIS as well as other workstations and peripherals within the hospital premises. It is these backbone closets that support this complex hospital network carrying critical data, voice, and video, keeping the network up and running. The hospital power system is a large complex electrical system consisting of high voltage transformers, automatic transfer switches (ATS), generators, isolation transformers, PDUs, etc. This power system feeds a variety of electrical loads including lighting, heating, ventila- tion air-conditioning (HVAC) systems, elevators, escalators, large pumps, fans, motors and more. The random nature of these loads (turning on and off randomly) creates an unstable power environment (i.e. sags and surges) that more sensitive imaging and diagnostic equipment and other IT networks that support them, must endure. Healthcare companies and hospitals faced with these challenges should engage partners like Schneider Electric with engineering expertise to perform complete physical infrastructure assessments that identify weaknesses and suggest corrective actions. This paper discusses and reviews the challenges imposed on physical infrastructure while deploying medical imaging and diagnos- tic equipment, with a focus on power and cooling. Depending on the patient’s ailment, a physician can use different modalities for diagnosis and treatment (i.e. X-Ray or CT for orthopedics, ECG or MRI for cardiology or Ultrasound for obstetrics). These modalities can be broadly classified into two categories: portable and stationary. Portables can be further classified into hand held (i.e. blood glucometer) and trolley or cart-mounted (i.e. ultrasound) while stationary devices can be further classified into desk-mounted (i.e. blood, urine analysis equipment) or floor-mounted (i.e. CT, MRI). Figure 2 illustrates a cart-mounted ultrasound and Figure 3 illustrates a floor-mounted MRI machine. The floor-mounted, desk-mounted, and cart-mounted devices need the most physical infrastructure planning. Modalities
  • 4. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 4 Environment Modalities are generally used in an indoor office environment. The cart-mounted and desk- mounted modalities generally use 120 VAC, 208 VAC or 230 VAC single phase power less than 5 kVA1 . The floor mounted devices typically require 208 VAC, 400 VAC or 480 VAC three-phase power, ranging from 20 kVA to 300 kVA or more. They require a lot of space and often times have their own separate room within the main hospital building or adjacent to it. They may be cooled with the building’s comfort air-conditioning system or they may have a precision computer room air-conditioning (CRAC) system, which more tightly controls temperature and humidity in the environment. Figure 4 illustrates a typical MRI facility with floor mounted equipment. Challenges Modalities need to be protected from power anomalies that cause hardware failures such as blown power supplies or printed circuit boards (PCBs) as well as from system software crashes. Physical space is a major constraint for large modalities like CTs and MRIs, more so in big hospitals in urban areas, as they have no room to expand. These modalities consume a lot of power, so heat dissipation is a major challenge to the building cooling system. Often times, comfort cooling is not sufficient and precision cooling is required. One of the most critical requirements is to provide electrical isolation from the electrical utility input to protect the patient and the technicians from any shock hazards. Compliance to local, state, national, and international codes is of paramount importance (i.e. National Electrical Codes (NEC) in the United States). 1 See White Paper 15, Watts and Volt-Amps: Powerful Confusion for information on kilo volt amps Figure 2 (left) Cart mounted ultrasound Figure 3 (right) Floor mounted MRI Figure 4 Typical CT or MRI facility
  • 5. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 5 Best practices • Since the hospital power grid is electrically “noisy and dirty” with a lot of electrical surges and sags, it is a good practice to provide UPS protection to all sensitive, expen- sive electronics systems, LCD displays, workstations, printers, and peripherals. The UPS system protects the hardware, avoids unwarranted system crashes while tests are in progress, prevents loss of patient data files, and provides safe, reliable radiology ex- aminations. • UPS systems used in the hospital should meet the following stringent standards: o UL1778 – American Standard for UPS o CSA22.2 No. 107.1 – Canadian Standard of UPS o FCC Part 15 Class A – American Standard for Electromagnetic Radiation o ANSI C62.41- American Standard for surge withstand capability o IEC60950 – International Standard for UPS Systems o EN50091-1- European Standard for Electromagnetic compatibility Depending on their usage, cart-mounted and desk-mounted devices may require the UPS to comply with the international standard IEC60601-1, “Standard for Medical Elec- trical Equipment, Part 1: General Requirements for Safety” for patient vicinity applica- tions. The U.S. equivalent is UL 60601-1. • For large floor-mounted modalities, a large UPS (50-300kVA typical) should be in- stalled to protect the entire room. Isolation transformers and the appropriate circuit breakers should be used to limit the leakage currents and electrical shock hazards for such modalities. These isolation transformers are sometimes built into the UPS or may be optional outside of the UPS box. All equipment should be installed in compliance with relevant codes like NEC, NFPA 70, NFPA99 (for U.S), and all other applicable lo- cal and national codes such as BS7671:2001 (U.K.), NFC15-100 (France), and VDE (Germany), as well as international such as CEI IEC 60364. Understanding and inter- preting the codes can, in some unique cases, become very controversial and the final say should be from the authorities having jurisdiction. In instances where budget con- straints prohibit investment in a large facility level UPS, a smaller UPS (5-10kVA) dedi- cated to the sensitive electronics and computer system of the CT, MRI, and PET should be considered. In addition, UPS protection should be provided to all their viewing sta- tions & workstations. • Sizing of the UPS for many devices like CTs and MRIs can be challenging since they draw very high amounts of inrush current. Ample precautions should be taken while sizing their power systems (including UPS, generators, transformers, and switchgear). Their normal power consumption as well as the inrush current ratings are available from their manufacturers. Allow enough margins for miscellaneous loads and future growth. Many companies like Schneider Electric have dedicated systems engineers, power protection specialists, etc. who can help evaluate the right solution for every unique customer situation. • Adequate cooling and air flow should be provided for all modalities that have sensitive electronics dissipating heat. For most of the cart and desk-mounted modalities, build- ing HVAC should be sufficient. However for large floor mounted modalities like CTs, MRIs, or PETs supplemental cooling may be required. Precision cooling is preferred as it can provide temperature and humidity control in the CT/MRI room. • All of the networked modalities and their physical infrastructure should be monitored and managed (i.e. environmental conditions of the radiology room, UPS battery life, runtime and capacity, and generator fuel) so that anomalies can be quickly detected and a corrective action be taken proactively to avoid any downtime.
  • 6. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 6 PACS make it possible to electronically store, manage, distribute, and view images. Funda- mentally, these systems are a network of all image acquisition devices, display workstations, and storage systems. They are made up of a broad range of technologies that enable digital radiology and digital hospitals to perform tele-radiology, tele-medicine, and tele-surgery. PACS have become more complex, encompassing systems that digitally acquire, convert, interpret, transmit, and store medical images. Diagnostic images will be available anytime, anywhere with little or no human intervention, making their distribution faster, easier, and more reliable. Figure 5 illustrates the components of a typical PACS. Environment The core of the PACS is made up of high availability RAID storage and server clusters running Windows, Unix, Linux, or a propriety operating system. These RAID storage and server clusters are housed in racks in a computer / data room or data center environment. Typically they draw less than 10 kVA, single-phase AC power at 120 V, 208V, or 230 VAC. Very large systems may draw three-phase power. Challenges PACS need to be available on demand to the nurse, physician, clinician, or specialist surgeon, providing latest imaging data of the patient under treatment. It needs to be highly available, 7x24x365 and there is little tolerance to downtime. Since the RAID drives and server clusters are confined to rack enclosures, handling their heat dissipation within the racks often becomes a bigger challenge. Best practices • PACS should be protected with an N+1 redundant UPS system. This N+1 UPS system not only protects the hardware but also protects the software from malfunctioning and gracefully shuts down and reboots the operating system if needed, thereby preventing a hard crash. The N+1 redundancy of the UPS system mirrors the redundancy of the RAID storage drives and server clusters which are at the core of the PACS, providing high availability. For smaller, simpler systems, a basic UPS can be provided. • Often times, additional receptacles are needed for plugging in all required devices. Rack based PDUs should be used to provide additional outlets. PDUs that can meas- ure and display current, which can help prevent accidental overloading and shut down of the PACS are recommended. PDUs that allow remote outlet control via the web are desirable as they can help reboot a hung server or a storage drive efficiently. Picture archiving and communication systems (PACS) Figure 5 Typical PACS
  • 7. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 7 • At a minimum, surge suppression should be provided for LCD/CRT based passive view-stations. For PC based active workstations running software applications, UPS protection with graceful shut down and reboot capabilities is highly recommended. • The PACS storage and servers should be housed in secured, lockable, rack enclo- sures. These racks should be in a temperature controlled environment. The racks housing PACS storage and servers are generally very dense physically and in terms of power consumption. The rack doors should be perforated, allowing for maximum air- flow. When power draw in the rack exceeds 4 kW, the rack should be placed in a hot or cold aisle containment system. If this isn’t possible, ducted racks should be used to capture hot return air and direct it back to the air conditioner. Alternatively, a row- based cooling unit can be used to provide supplemental cooling capacity (see Figure 6). For more information on containment systems for existing data centers see White Paper 153, Implementing Hot Aisle and Cold Aisle Containment in Existing Data Cen- ters. • A good management strategy involves the management of PACS servers, storage, and their entire physical infrastructure including UPS, PDUs, batteries, and their critical en- vironment (temperature and humidity). This will give early warning of any anomaly or impending disaster so that corrective actions can be taken and prevent all avoidable shutdowns. RIS and HIS are server based systems running special software that make it possible to store, monitor, manage, and distribute patient medical information. They help patients in scheduling appointments, registration, and billing, and help hospitals in generating, maintain- ing, and managing patient’s electronic medical records as well as generate workflow, work- list, management reporting, and variety of other tasks. These RIS and HIS are really becoming one large HIS and are integrated / networked with PACS as well as various other modalities within the hospitals providing complete automation. By converting them into “digital hospitals”, they can significantly improve patient-care, minimizing human errors, saving lives, and reducing costs. Figure 7 shows a typical RIS / HIS and its subcomponents. Environment These systems are generally housed in a data center environment drawing 10 kW single- phase 208 VAC or 230 VAC power on the lower side, to hundreds of kilowatts of three-phase 400 VAC or 480 VAC power on the higher side. The majority of data centers in hospitals have a UPS with battery back-up, precision air conditioning units, and a back-up generator. Challenges RIS / HIS are the most important systems within the data center requiring longer runtime and higher redundancy and availability then most other equipment. Since these systems are Figure 6 Example of ducted rack air containment (left) (Schneider Electric Vertical Exhaust Duct shown) Example of row-based cooling unit (right) (Schneider Electric InRow RC shown) Radiology information systems (RIS) and hospital information systems (HIS)
  • 8. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 8 merging to form one big HIS on which the entire hospital depends for normal functioning, their availability requirements are generally 99.999% (five nines) or higher which translates to average unplanned downtime of 5 minutes per year or less. Additionally these systems may be located in high rise buildings and attention should be paid to the floor load (weight) handling capacity, elevator hauling capacity, door heights and widths to ensure that the physical infrastructure elements like UPS, batteries, and air-conditioning can be rolled in to their planned positions. Hospital Wards Transcriptionists Reception Management Reporting Medical Technical Department Clinical Departments Best practices • The physical infrastructure supporting the RIS / HIS should provide highest levels of redundancy while minimizing the total cost of ownership. An N+1 redundant UPS with automatic and manual bypass is very common and often times it is extended to the generator as well as the precision air-conditioning systems to ensure the highest levels of availability. The entire infrastructure should be scalable to allow for future expan- sion, be manageable like the other IT equipment, and be serviceable to reduce mean time to recover. An example of such a system is the Schneider Electric InfraStruxure and is shown in Figures 8. All of these characteristics contribute to the overall availa- bility of the system. • Servers and systems requiring the highest levels of availability should be identified and grouped so that they can be provided with longer runtime and higher levels of redun- dancy in a separate area, and in separate racks within the data center. This concept of “targeted availability” helps increase availability of business critical systems without having to incur a large capital expense for the entire data center. Higher levels of re- dundancy like dual feeds with dual generators and dual N+1 UPS with dual power paths all the way to the rack should be considered for highly-critical data centers and networks. • PDUs should be able to measure and display current, which can help prevent acci- dental overloading and shutdown of the RIS / HIS. PDUs that allow remote outlet con- Figure 7 Typical RIS / HIS system Figure 8 Schneider Electric InfraStruxure™
  • 9. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 9 trol via the web are desirable for rapidly rebooting a hung server or a storage drive. Isolation transformers should be used wherever required and mandated by local laws. • Precision air conditioning equipment should have the capability to allow for expansion. Redundant air conditioning units should be considered for higher availability. For high power density racks (>4 kW / rack) additional air distribution and air containment should be used to avoid hot spots. For more information on cooling best practices refer to White Paper 49, Avoidable Mistakes that Compromise Cooling Performance in Data Centers and Network Rooms. Medical imaging and diagnostic equipment are typically connected to a network. Modalities like CTs and MRIs get connected to PACS which are connected to RIS and HIS which in turn are connected to the hospital intranets and extranets. The wiring closets or IDFs, as shown in Figure 9, play a very vital role in ensuring the connectivity of this equipment and the availability of the network, 7x24x365. Wiring closets comprise of layer 2 and layer 3 access and distribution switches, hubs, routers, patch panels, UPS systems with a battery back-up as well as any other miscellaneous telecommunications equipment mounted generally in a two post rack. IDFs or wiring closets may also supply power over Ethernet (PoE) to net- worked devices likes IP phones, web/security camera and any other devices drawing power up to 15 W. This imposes a lot more challenges on the power and cooling requirements in the closets. Environment These IDFs or wiring closets are typically hidden in some remote location of the building with little or no ventilation and illumination. Legacy telecommunication networks typically used wiring closets mainly for punch-down blocks, patch panels, and a few small stackable hubs or switches. However, networking equipment that supplies power over Ethernet uses and dissipates considerably more power. These new switches support data, voice, and video, generally 19” rack mount type, and have varying air flow patterns depending on the manufac- turers (i.e. side to side vs. front to back). A typical IDF will house 1-3 racks worth of equip- ment and draw 500 W to 4000 W of single phase AC power or more. Two post racks have largely been replaced with four post racks since newer equipment is getting heavier and deeper. Wiring closets or intermediate distribution frame (IDF) Patch Panel Midspan Power Supply Network Telephony System Network Switches Uninterrutible Power Supply Figure 9 IDF (wiring closet)
  • 10. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 10 Challenges While deploying PACS, RIS, HIS, or new modalities that are being networked, these wiring closets or IDFs need the most attention in terms of power and cooling. It can be a challenge to ensure the right type of receptacles (i.e. L5-20, L5-30, L6-20, IEC 320 C19, and IEC 320 C13) and the right amount of power with the right circuit breaker protection to all the network- ing equipment. Cooling and airflow are often a bigger but ignored problem to address in these wiring closets. Best practices All equipment in the IDF should be protected by a UPS system. The selection of UPS should be based on: • The total power required in Watts • The run time required in minutes • The level of redundancy or fault tolerance desired • The voltages and receptacles required The UPS system is sized by taking the sum of the Watt ratings of the loads. A common rack- mount UPS (i.e. Smart-UPS™) will provide approximately four nines (99.99%) of power availability, while an N+1 redundant, UPS with built in bypass (i.e. Symmetra RM) with one hour runtime will provide approximately five nines (99.999%), which may be sufficient for most applications. See Appendix of White Paper 69, Power and Cooling for VoIP and IP Telephony Applications for details on this availability analysis. UPS products are available with battery packs to provide different durations of run time. Identify the plugs and receptacles required for all the equipment including the UPS in the wiring closet. Ideally all of the equipment should be plugged directly into the back of the UPS or the transformer, and the use of additional outlet strips or rack PDUs should be avoided. However, if there are many devices, it may not be practical and a rack PDU should be used. In that case a high-grade rack PDU specifically designed for the purpose should be used. The PDU should have enough receptacles to plug all the current equipment with some spares for future needs. PDUs with a meter displaying the current power consumption are preferred as they reduce human error like accidental overloading and resultant load drops. For the correct selection of the appropriate UPS model meeting the required power level, redundan- cy, voltage, and run time, the process is simplified by using a UPS selector such as the APC by Schneider Electric UPS Selector. This system has power data for all popular switches, servers, and storage devices, which avoids the need to collect this data. In systems like this, the choice of configuring a UPS will provide various receptacle options. To ensure continuous operations of the equipment in the wiring closet, 7x24x365, cooling and airflow issues must be identified and addressed. The problem of heat dissipation and need for supplemental air-conditioning is most pronounced in the wiring closets which have no vents. Power dissipation in the wiring closet should be calculated to decide on a cost effective way to solve the problem (see Table 1 & Table 2 in White Paper 69, Power and Cooling for VoIP and IP Telephony Applications for details). Finally, environmental monitoring (i.e. temperature and humidity) within these wiring closets is highly recommended as it will help flag any abnormal conditions, allow for enough time to take proactive measures, and avoid downtime.
  • 11. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 11 To ensure high availability and reliability to medical imaging and diagnostic equipment, including PACS, RIS, HIS, modalities, and their network, special attention must be paid to their physical infrastructure. The biggest challenges are in terms of power, cooling, physical space, management, and services. Providing UPS protection to all such devices protects the hardware, prevents the software from unwarranted crashes, and increases their availability significantly. Cooling is a special problem for bigger floor mounted modalities, high density storage and servers for PACS as well as RIS / HIS and hospital wiring closets. In some cases, a building’s HVAC system along with proper ducting, ventilation, and airflow may be sufficient. However in many situations, additional cooling in the form of precision air condi- tioning is required. Companies like Schneider Electric have dedicated team of systems engineers, power protection specialists, and availability consultants who can help in doing assessments and audits of data center physical infrastructure and provide detailed actionable reports on improving overall system reliability and availability while minimizing the total cost of ownership. Conclusion Special thanks to Viswas Purani for authoring the original content of this white paper. Acknowledgements
  • 12. Power Protection for Digital Medical Imaging and Diagnostic Equipment Schneider Electric – Data Center Science Center Rev 0 12 The Different Types of UPS Systems White Paper 1 Avoiding costs from Oversizing Data Center and Network Room Infrastructure White Paper 37 Dynamic Power Variations in Data Centers and Network Rooms White Paper 43 Avoidable Mistakes that Compromise Cooling Performance in Data Centers and Network Rooms White Paper 49 Power and Cooling for VoIP and IP Telephony Applications White Paper 69 Implementing Hot Aisle and Cold Aisle Containment in Existing Data Centers White Paper 153 IEEE White book “IEEE Recommended Practice for Electric Systems in Health Care Facili- ties” IEEE std. 602-1996 References Resources Browse all white papers whitepapers.apc.com tools.apc.com Browse all TradeOff Tools™ ©2014SchneiderElectric.Allrightsreserved. For feedback and comments about the content of this white paper: Data Center Science Center dcsc@schneider-electric.com If you are a customer and have questions specific to your data center project: Contact your Schneider Electric representative at www.apc.com/support/contact/index.cfm Contact us