2. Overview
1
Removable
air pathway
Remote
alarm
connector
Ethernet
connector
Serial connector – to
connect to a computer or
Alice polysomnographs
Air inlet and grey
foam filter
External
battery
connector
Detachable
battery
Oxygen valve port
(flow source up to
15 LPM)
Exit
cm
H2O
Primary: S/T AVAPS
0 5 10 15 20 25 30
24/04/2007
04:53 PM
Navigate Select
7
1 2 3
Cancel action or exit menu
Navigate menus or change settings
Select/modify/confirm parameters
Audio pause button
Start/stop button
AC power inlet
1
2
3
4
5
6
7 SD card to download ventilation data
4
Prescriber environment, full access mode
SD card inserted in the ventilator
Audio pause active
Ramp active
Status of detachable battery
Status of internal battery
Status of external battery
Battery is charging
Symbols and icons
5
6
3. 2 Circuit set-up
Before operating Trilogy100, you need to configure the circuit type.
Set-up with exhalation valve Set-up with exhalation port
Tee connector
Proximal
pressure line
Mask without
integrated leak
Exhalation
valve line
Flexible
tubing
Tracheostomy
tubing
Proximal
pressure port
Exhalation valve
Bacterial/viral filter
(optional)
Flexible
tubing
Whisper
Swivel II
(exhalation
port)
Whisper
Swivel II
(exhalation
port)
Mask with
integrated leak
Mask without
integrated leak
Tracheostomy
tubing
Bacterial/viral filter
(optional)
To change the
porting block Proximal pressure line
Exhalation valve line
Please check that circuit type is properly configured.
4. 3 Set-up screen
Setting the Active or Passive circuit
Make sure Trilogy100 is turned off before taking the following steps
Menu 1/10
Settings And Alarms
Options
Alarm log
Event log
Information
Exit
Navigate Select
Select SettingsAndAlarms Menu Settings and Alarms 2/23
Dual Prescription OFF
Circuit Type Active PAP
Mode S/T
Cancel
Edit OK
Use to choose Active PAP (exhalation
valve set-up) or Passive (exhalation port set-up)
and then press twice to confirm
B D
Menu Settings and Alarms 2/23
Dual Prescription OFF
Circuit Type Passive
Mode S/T
Exit
Navigate Modify
A C
Press and hold and for a few seconds to access Set-up
page
Navigate to CircuitType and press
5. A
Settings And Alarms
Options
Alarm log
Event log
Information
Menu Options 1/18
Menu Access Limited
Detailed View OFF
Language English
Navigate to Options and press Press to enter MenuAccess
C
E
Menu Options 2/18
Menu Access Full
Detailed View OFF
Language English
Press to begin therapy or
continue navigating through the menu
to set-up the ventilation parameters
To enable full access to the settings
B
D
Menu Options 1/18
Menu Access Full
Detailed View OFF
Language English
Use to choose Full or LimitedAccess
and press to confirm
Press and
to return to Initial Screen
Exit
Navigate Select Finish
Navigate Modify
Cancel
Edit OK Finish
Navigate Modify
Setting menu access
6. Settings And Alarms
Options
Alarm log
Exit
Navigate Select
Select SettingsAndAlarms
4 Prescription set-up
B
Menu
A
From main screen, press Menu
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
Dual Prescription
Menu Settings And Alarms 1/20
Dual Prescription ON
Mode S/T
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
Finish Navigate Modify
If Dual Prescription is enabled, both prescriptions
need to be adjusted as previously explained
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
RR 20BPM
PIP 5.9cm
H2O
Vte 435ml
Leak 21l/min
Menu 3/8
Switch to Secondary Settings
Safely Remove SD Card
Primary Settings and Alarms
Secondary Settings and Alarms
Options
Exit
Navigate Select
F
Two prescriptions to set
Peak Flow 67.9l/min
MinVent 8.7l/min
Cancel
Edit OK
Press to validate new setting
Cancel
Edit OK
E Use to change the value or status
of setting
7. AVAPS OFF
IPAP 30.0 cmH20
EPAP 8.0 cmH20
Finish
Mode S/T
AVAPS OFF
IPAP 30.0 cmH20
EPAP 8.0 cmH20
Navigate Modify Finish
Navigate Modify
Press to change the parameter
highlighted in blue
C D
Scroll up and down using to navigate
through the settings and alarms
Mode S/T
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
Finish Navigate Modify
G
Press and
to return to the main screen